18
X-Cite
®
Optical Power Measurement System User Guide
035-00390R rev. 3
Favorite Wavelength:
If already entered into the XR2100, the favorite wavelengths will self-populate.
Type in favorite wavelengths in each box, (will reflect in the power meter display).
To clear individual favorites
enter ‘
0
’ into each box, or clear all favorite
wavelengths from the
‘Power Meter’
pull-down menu.
Wavelength range: 320nm to 750nm.
8.5 Status Data
This window will self-populate once the XR2100 is connected.
Note:
S/N
= the XR2100 serial number
Items That Can Be Changed By Operator;
LCD Contrast:
Can be set between 1-15 (boldest LCD contrast to dimmest). Constantly
operating on the highest LCD contrast will shorten battery life.
Log during CAL:
Saves values to data log during a calibration.
Disable Rel Abs:
Locks the “Relative/Absolute” button on the XR2100. XR2100 stays in current
display mode.