16.4 Air Vents & Filters
X-Cite units have several air vents that are an integral part of the cooling system
and overall lamp performance. Proper cooling ensures that lamps operate at
optimal temperature and pressure for output power, spectrum, lamp life, light
guide life and safety.
Ensuring adequate cooling
1. Never obstruct the air vents on the X-Cite unit. Vents are located at the rear and
underside of the unit. (Note: Arrows indicate direction/location of air flow. In
earlier X-Cite models vents were located on the side, underneath the lamp
access panel.)
2. Always leave clearance for air flow between the X-Cite and walls or other
equipment. Do not remove the rubber feet on the X-Cite unit or otherwise
reduce/block the space between the bottom of the unit and bench top. This may
compromise airflow through the unit.