18
X-Cite
®
120PC Q User Guide
035-00328R rev. 4
7
X-Cite
®
120PC Q Message Reference
Symbol
Message
Description
XXXX.
Lamp Hours
Displays the accumulated hours the lamp has
been on
XXX.X
Exposure Time
Displays the time in seconds the shutter will
remain open when activated.
XXX
Iris Setting
Displays the iris setting in a percentage of
maximum.
Flashing
“
XXX.X
”
Flashing display
The lamp is warming up. Warm up time is
approximately 90 sec.
bulb
Bulb Error
Lamp installed incorrectly /
Lamp did not strike or extinguished after
striking
cool
Cool Warning
Lamp is too hot to strike.
The lamp will automatically strike when it has
reached optimum striking temperature
old / bulb
Alternating Old and then Bulb
The lamp has accumulated over 2500 hours.
Lamp may be near end of life.
end / bulb
Alternating End and then Bulb
The lamp has accumulated over 4000 hours.
The lamp will not strike.
LOC
Adjustment Locked
The
UP
/
DOWN
adjustment buttons have been
locked. No changes can be made to the
exposure time or iris setting.
ULOC
Adjustment Unlocked
The
UP
/
DOWN
adjustment buttons have been
unlocked. Changes can be made to the
exposure time and iris setting
SFI
Shutter Failure
The shutter has failed. Unit should be
restarted. If the error repeats unit should be
serviced.
See Section 16 for troubleshooting.