31
X-Cite
®
120PC Q User Guide
035-00328R rev. 4
15 Clearing an Audible Alarm
15.1.1 The X-Cite
®
120PC Q provides an audible alarm to alert the user of various error
conditions. The following conditions will cause the unit to indicate an alarm
condition:
i)
Lamp fail to strike
ii)
Lamp access door not correctly installed
iii)
Intelli-lamp sensor/ connection failure
iv)
Shutter position failure
15.1.2 To clear the audible alarm:
1. Press the
START
/
STOP
button
16 Troubleshooting
16.1 General
Service to be completed by qualified repair personnel only!
If the unit fails to power up, use the following checklist to eliminate the most
common causes of problems.
Check that:
1. The AC power cord is securely plugged into a functional AC wall plug.
2. The AC power cord is securely plugged into the AC inlet on the rear of the unit.
3. The mains AC power switch is in the ON position.
4. Check that the ventilation openings on both the bottom and rear of the unit are
not blocked.
If the unit still does not power-up:
1. Check both mains power fuses by first disconnecting the power cord. Then
carefully remove the fuse drawer assembly below the AC inlet on the rear of the
unit. If the fuse(s) is/ are open, replace with the same type (4A, 250 V, Fast
acting).