33
X-Cite
®
120PC Q User Guide
035-00328R rev. 4
If the LED display indicates “SFI”:
1. This indicates there is a shutter failure.
2. Power down the unit wait 1 minute and turn the power back on.
3. If the unit once again displays the shutter failure message,
If the LED display does not light:
1. If the fan is functional, power down the unit, wait approximately 20 seconds then
power it up again
contact your local X-Cite Service Center.
If either Fan does not work properly
1. If the LED display is functional, power down the unit, wait approximately 20
seconds then power it up again
contact your local X-Cite Service Center
16.2 Liquid Light Guides
Proper handling and installation of liquid light guides
1. Liquid light guides have a typical useful life of 4000 hours of operation when
handled properly and installed in a well maintained X-Cite 120. The formation of
bubbles is one of the most common reasons for a light guide to degrade
prematurely and result in a sudden reduction in illumination intensity. Bubbles
can form without warning, and if they occur within the first 1500 to 2000 hours of
use, is it typically due to overheating and/or mechanical stress to the light guide.
Below are some simple tips to avoid overheating and stressing the light guide.
2. Always fully insert the light guide in to the X-Cite unit (i.e.
up to the “A” line
printed on the grey ferrule); this ensures contact with a heat sink to conduct heat
away from the light guide.