background image

 

10 

Cartridge Installation and Activation 

 
1.  Remove  the  outer  shrink-wrap  and  protective  closure  from  the  pre-filter,  TO  and  RO 

cartridges. 

 
2.  Remove the sump from the pre-filter housing. Install the pre-filter into the sump. Locate the 

O-ring  in the  small  parts  pack.  Lightly  lubricate  this  O-ring  with  a food grade  lubricant  and 
install it into the groove  in the sump. Re-install the sump with pre-filter and O-ring into the 
pre-filter head. 

 

3.  Open the TO flush valve. Install the TO cartridge into the TO head (right hand), following the 

instructions on the cartridge. Position a bucket or other container under the TO flush valve. 
Slowly open the inlet shut-off valve. Carbon fines and air will be flushed out of the TO flush 
valve  during  this  time.  Flush  until  water  runs  clear  (3-5  minutes  typically).  Close  the  inlet 
shut-off valve and the TO flush valve. 

 
4.  Install the RO cartridge into the RO head (left hand). Rotate the handle on the product water 

3-way sample valve so it is pointing towards the drain connection (front of unit). Close the 
filtered water blend valve. 

 
5.  Slowly open the inlet shut-off valve. Position a bucket or other container under the product 

water 3-way sample valve drain connection. Slowly open the inlet shut-off valve. Plug the 24 
VAC transformer into a 115-Volt receptacle. Water will begin to flow out of the RO discharge 
tubing and product water 3-way sample valve drain connection. 

 

WARNING:

  Do not use this product water! It may contain a preservative solution from the 

RO cartridge and should not be consumed! 

 
6.  If  you  are  utilizing  the  capillary  tubing  assembly,  refer  to  the  section  titled  “Adjusting  the 

Recovery” for instructions. Return to this section when complete. 

 
7.  Allow  the  system  to  flush  for  at  least  1  hour  or  until  1  gallon  of  product  water  has  been 

produced. Close the inlet shut-off valve. 

 

Start-Up and Blend Adjustment 

 
1.  When flushing is complete, the storage tank requires filling with chlorinated water, time for 

effective  disinfection,  and  then  complete  draining.  The  steps  for  this  are  outlined  below. 
Prepare for draining the storage tank by doing one of the following: 

 

a)  Remove the drain plug on the storage tank and temporarily install a nipple, valve and 

hose or tubing to drain. 

b)  Disconnect the tubing/piping connected to one of the systems being serviced by the 

RO. Divert the tubing/piping to drain as required. 

 
2.  Rotate  the  handle  on  the  product  water  3-way  sample  valve  so  it  is  pointing  towards  the 

storage tank connection (rear of unit). The system will now begin to fill the storage tank with 
product water. Remove the storage tank access cover. Pour 2 ounces of 5.25% bleach into 
the storage tank. Open the filtered water blend valve fully and allow the storage tank to fill 
until the level switch stops the flow of water. Close the filtered water blend valve. Disconnect 
the 24 VAC transformer from the 115-Volt receptacle. Manually add additional water to the 
storage  tank  until  completely  full.  Reinstall  the  storage  tank  access  cover.  Allow  the 
chlorinated water to remain in the storage tank for 10 minutes to disinfect the storage tank 
effectively. Close the filtered water blend valve. 

 

Summary of Contents for MRS-200

Page 1: ...1 Cat No EV311117 Rev B 11 6 03 DCO 7060 MODEL MRS 200 Reverse Osmosis Mineral Reduction System Installation and Start up Guide...

Page 2: ...Installation 4 Cartridge Installation and Activation 10 Start Up and Blend Adjustment 10 Adjusting the Recovery 12 Operation 14 Maintenance 14 Performance Monitoring 16 TDS Measurement 16 Performance...

Page 3: ...hlorine An activated carbon filter has been provided with this system to protect the reverse osmosis membrane from chlorine attack Influent chlorine should not exceed 3 mg L The MRS 200 System consist...

Page 4: ...nd permit access for maintenance The system will weigh approximately 175 pounds when the storage tank is full 4 ____ Does the inlet water supply meet the requirements listed below 1 A minimum of 1 0 G...

Page 5: ...nts Place the main unit on the shelf positioning the cartridge heads directly over the cutout area Insert a cable tie into one of the holes in the baseplate and through the wire slats in the shelf mak...

Page 6: ...A MP S 2 INS E RT E LB OW FULLY 3 S LIDE LOCK ING CLIP UP TO S E CURE 1 S LIDE LOCK ING CLIP DOW N Figure 3 4 Water supplies with a dynamic operating pressure of 50 PSI or less require the use of the...

Page 7: ...UMP 10 NUT 8 1 2 X 1 SPACER 4 10 X 1 1 2 SCREW 8 10 WASHER 16 MOUNTING PLATE Figure 4 5 Refer to Figure 5 for a sample general arrangement Use this as a guideline for determining valve location tubing...

Page 8: ...ow control insert red plastic cylinder in small parts kit Press this insert into the red OD tubing Fig 6 Attach the end of the tubing with the flow control insert into the 3 8 stem x OD tube reducing...

Page 9: ...lt receptacle at this time Figure 9 Figure 7 Figure 8 INLET WATER SUPPLY 50 PSIG MAX TO 120 VAC OUTLET A B C SOLENOID VALVE 1 5 2 4 3 IN A B C TO 120 VAC OUTLET SOLENOID VALVE IN COVER REMOVED 3 4 2 5...

Page 10: ...ative solution from the RO cartridge and should not be consumed 6 If you are utilizing the capillary tubing assembly refer to the section titled Adjusting the Recovery for instructions Return to this...

Page 11: ...to contain a range of TDS greater than what an RO system normally produces In those cases a filtered water blend valve has been provided to blend in filtered not RO treated water to attain a specific...

Page 12: ...product flow with a 1000 ml graduated cylinder Don t worry if the flow is slightly more or less than what is specified 2 Determine the appropriate concentrate flow for your system and the desired rec...

Page 13: ...480 70 520 80 554 10 00 50 408 60 450 70 489 80 521 12 00 50 363 60 402 70 437 80 467 14 00 50 330 60 366 70 399 80 427 16 00 50 307 60 342 70 371 80 397 18 00 50 292 60 326 70 352 80 377 20 00 50 28...

Page 14: ...or 6 000 12 000 18 000 or 24 000 gallons d Calendar scale is jumper selectable for 6 months or 12 months Monitor operation a Active green light blinks every two seconds during any minimum water flow c...

Page 15: ...s or until the water is clear 10 Close the TO flush valve 11 Connect the 24 VAC Transformer from the 115 Volt receptacle Every 6 12 months Sanitize the storage tank 1 Refer to Start up section steps 1...

Page 16: ...and a graduated cylinder or calibrated measuring cup The drain water flow rate can be measured from the end of the red tubing in a similar manner Inlet water flow Outlet Product Water Flow Drain Wate...

Page 17: ...has removed 92 of the dissolved solids from the inlet water 6 The RO cartridge should be changed every 1 to 1 1 2 years or when the percent reduction falls below 90 A form has been provided below to r...

Page 18: ...valve as required to attain specified water quality a Tubing piping run to equipment too restrictive a Increase tubing piping diameter Refer to System Installation section for recommendations 4 Low w...

Page 19: ...19 Replacement Parts...

Page 20: ...EV307850 Support Ring EA 2 49 EV310445 Label Carbon Filter EA 1 50 EV310446 Label RO Module EA 1 51 01010248 Adaptor 3 8 Stem X 1 4 Tube EA 1 53 EV306996 Directional Arrow EA 1 No Part Number Descrip...

Page 21: ...hole punched Prints black ink on 50 offset white VENDOR MUST SUBMIT A MATERIAL CERTIFICATE OF COMPLIANCE WITH EACH SHIPMENT LET CHANGE BY APRVD DATE A DCO 4785 RH SK 9 12 02 B DCO 7060 KR RH 11 6 03 T...

Reviews: