Ensure
correct delivery of medication
•
Before each injection, carry out a function check (Chapter “Priming / Function Check” step 1 – 4) to prevent
incorrect dosage caused by air in the cartridge or a blocked pen needle. Use a new pen needle if priming is not
successful.
•
Check after each injection according to your physician’s and/or health care professional’s instructions whether
medication drips from the pen needle or D-
mine
®
Pen, or exudes from the injection site. If the medication drips
from the pen needle or D-
mine
®
Pen, or exudes from the injection site, follow your physician’s and/or health care
professional’s instructions regarding the handling of potential under-dosage. Seek medical attention if you have
any health concern.
•
In the event of interruption of injection by releasing the push-button, check on the display the amount of medicine
not delivered. The remaining amount can be injected by pressing the push-button again.
•
In the event of accidental medication injection seek immediate medical attention.
CONFIRMED LIFE SPAN
The continued dosing accuracy and requirements according to EN ISO 11608-1 have been verified for the amount of injec-
tions and cartridge exchanges as defined hereinafter in the table below from the date of first use:
Daily dose
[mg]
Injections/
day [#]
Dose/injecti-
on [mg]
Life span
[months]
Daily dose
[mg]
Injections/
day [#]
Dose/injecti-
on [mg]
Life span
[months]
3
3
1
36
30
6
5
18
6
3
2
36
36
6
6
18
9
3
3
36
35
7
5
12
12
3
4
36
42
7
6
12
15
3
5
36
40
8
5
12
16
4
4
30
48
8
6
12
20
4
5
30
45
9
5
12
24
4
6
30
54
9
6
12
25
5
5
24
50
10
5
12
30
5
6
24
60
10
6
12