100 MM GRAPHICS RECORDER: USER GUIDE
User Guide
Page 39
HA027271
Issue 12 Mar 04
4.2.5 Import User Linearisation
The Import Linearisation feature allows linearisation tables to be imported either via the floppy disk, or if the Remote
Viewer option is fitted,directly from the host PC.
Touching the User Linearisation field allows the user to select which of 'UserLin1' to 'UserLin4' is to contain the im-
ported file.
Touching the File Name field, calls a popup menu similar to that shown in
. This allows the user to select
the table to be imported.
The file must be comma separated ASCII as follows:
n,
X1,Y1
X2.Y2
X3,Y3
.
.
Xn,Yn
where 'n' is the total number of XY pairs in the table, and or each pair, X is the input value, and Y is the linearised
value corresponding to X.
See
Note: Imported linearisation tables will not become effective until after the next configuration 'Apply'
operation
4.2.6 Export User Linearisation
Similar to 'Import user Linearisation, above, the Export Linearisation feature allows linearisation tables created in the
recorder to be exported either via the floppy disk, or if the Remote Viewer option is fitted,directly to the host PC.
See
Note: If an exported linearisation table is to be used in a recorder with software version prior to version 2.2, it
must be exported in Numeric format (i.e. not in Scientific format).