Remote Operation
|
51
—Read Battery Voltages
Returns the current battery percentages of probes in use.
Syntax:
RBAT?
Parameters:
None
To check the current battery percent of the probes in use:
RBAT?<LF>
Response:
62%,100%,,<CR><LF>
Channels 3 and 4 are not on or not ready.
—Read Peak
Writes the peak probe field values to a remote port.
Syntax:
RDPK,x
Parameters:
(x)
1 = Reads channel 1 max
5 = Reads channel 1 min
2 = Reads channel 2 max
6 = Reads channel 2 min
3 = Reads channel 3 max
7 = Reads channel 3 min
4 = Reads channel 4 max
8 = Reads channel 4 min
Output format:
Ch1 Max xx.xx,Xxx.xx,Yxx.xx,Zxx.xx<CR><LF>
—Reset Peak
Resets peak probe field values.
Syntax:
RSPK
Parameters:
None
Summary of Contents for HI-6100
Page 1: ...HI 6100 Field Monitor User Manual HI 6100 Field Monitor shown with optional probes ...
Page 12: ...12 Introduction This page intentionally left blank ...
Page 16: ...16 Maintenance This page intentionally left blank ...
Page 18: ...18 Specifications This page intentionally left blank ...
Page 24: ...24 Assembly and Installation This page intentionally left blank ...
Page 27: ...Operation 27 Overview of HI 6100 Menu Options Figure 7 Overview of HI 6100 Menu Options ...
Page 56: ...56 Remote Operation This page intentionally left blank ...
Page 58: ...58 Error Handling and Troubleshooting This page intentionally left blank ...
Page 60: ...60 Warranty This page intentionally left blank ...
Page 61: ...EC Declaration of Conformity 61 Appendix B EC Declaration of Conformity ...
Page 62: ...62 EC Declaration of Conformity This page intentionally left blank ...
Page 70: ...70 Commands in FM5004 Mode This page intentionally left blank ...