46
Molift QuickRaiser 205 / www.etac.com
Periodic Inspection Checklist QuickRaiser 205
In accordance with ISO:10535 2006 Annex B.
Article no.:
...........................................................................................
Serial no.:
...........................................................................................
Production Year:
...........................................................................................
Owner:
...........................................................................................
Sling shall be inspected separately and is not
included in this inspection.
Situation of use:
Home
Institution (type of institution):
Hospital
Nursing home
Other
Service light indicator
Status
Service light Mode
No light
No service needed
Green
flashing
Order service
Yellow
Periodic inspection
required
Red
Perform service
Physical examination
Test FUNCTION and check for wear.
All checkpoints must be checked of
to approve lifter for further use.
Perform 1 lifting cycle with the weight of
a person (minimum 50kg/110lbs). Brake
shall not slip, and the lifter shall have no
abnormal sounds.
Emergency stop, activate and reset switch.
Hand control (all buttons).
Control panel (all buttons, excluding electri-
cal emergency lowering).
Check tools for emergency lowering
Leg opening mechanism.
Brake on castors.
Charger and battery
Accessories:
...........................................................................................
Visual examination
Visual examination of load bear-
ing structure to make sure there
is no damage, cracks, frays or
deformation. All checkpoints must
be checked of to approve lifter for
further use.
Product labels
Plastic covers
Hand control and control panel
Cables and other electronics.
Rubber profile (on column)
Leg opening mechanism
Lifting arm and composite hooks
Lifting arm connection (remove cover)
Visible weldments on welded parts
Performed by
Periodic inspection should be performed by a
person who is certified by Etac education.
Date/Place:
...........................................................................................
Signature:
...........................................................................................
Etac Molift Service ID (6 digits):
...........................................................................................
Approved without faults, service indicator
has been reset.
Approved after repair, service indicator has
been reset.
Lifter is marked with “out of order” and
sent for repair.
Hoist is not eligible for repair and taken out
of service.
If periodic inspection reveals any defect, wear or
other damage that jeopardizes the safety of the
patient it may not be used until the deficiency has
been eliminated. Defects and damages should be
reported back to the manufacturer for action in a
non-conformity report (NCR).
The owner is notified
NCR report sent to Etac, molift.ncr@etac.
com
Non conformity report can be obtained from
www.etac.com or by request, [email protected].
Summary of Contents for Molift QuickRaiser 205
Page 1: ...TM13001 Rev B 2018 01 03 Molift QuickRaiser 205 EN Technical manual...
Page 12: ...11 Molift QuickRaiser 205 www etac com Driveline Article no 2910030 6 6 1 4 5 3 2...
Page 50: ...49 Molift QuickRaiser 205 www etac com Manufacturer s Declaration...
Page 51: ...50 Molift QuickRaiser 205 www etac com Manufacturer s Declaration...
Page 52: ...51 Molift QuickRaiser 205 www etac com Manufacturer s Declaration...