Troubleshooting
Recycling
Recycling instructions are available at Etac.com
Symptom
Possible Cause/Action
The sling does not
fit the user.
Wrong sling size. Try another size.
The sling is too small when the leg sup-
ports do not reach halfway around the
inner thigh and the suspension is too close
to the user’s face. Try another size.
The sling is not correctly positioned.
Pull the sling further down at the back
when fitting it.
All the straps do
not lift at the
same time.
If the sling straps lift unevenly, it means
that the sling can slide up at the back
which can lead to the user sliding down
into the opening of the sling. Make sure
that the leg supports are positioned cor-
rectly on the thighs.
User is sliding
through/out of
the sling.
The sling is not connected correctly
according to the users individual
prescription.
The user is wearing to slippery clothes.
Change or remove some of the clothing.
The sling is too big, try a smaller size or
another sling
1. Fold the SlingOn in a way that makes it easy to slide and pull
underneath the patient. The SlingOn must be doubble-layered
with an opening at the top of the bed, this will make the applica-
tion of the sling more efficient.
2. Slide the SlingOn under the patients head, a caregiver on each
side of the patient should pull the SlingOn simultaneously down
until it is fully unfolded and covering the patients body.
3. Slide the sling between the two layers of the SlingOn, starting
from the top and working your way down.
5. Make sure the lifting straps are
the same length on both sides and
that the sling is placed under the
users thighs without any creases.
Depending on the patients needs,
the central legstraps can be crossed
or not. If crossed, pull pne of the
loops through the other. Connect all
6 lifting straps to the suspension.
4. Position the sling anatomically correct underneath the patient
before removing the top layer of the SlingOn. Raise the back
of the bed to make the patient have a sitting position before
performing the lift.
We suggest the use of the Immedia SlingOn
to apply the MediumBack PLUS in a correct
and safe manner.
Remove the sling in reverse order
Not crossed
Crossing
Crossed