4. The equipment or system should not be used next to or placed on top of another device. If use in the
vicinity of another device or with the device installed on top of another device is required, it must be
checked for proper operation in the exact configuration where it will be used.
5. Essential performances : Product integrity is as described on the leaflet after immunity tests. The
device is designed according to ISO 10341
6. Appliance shall fulfill these essential performances for all Immunity tests (ESD, Radiated RF
disturbance, Fast transient, Surge, Conducted RF disturbance, Magnetic field).
Manufacturer guidelines and declaration - Electromagnetic emissions
Vision-R
TM
800 is intended to be used in the electromagnetic environment specified below.
It is up to the customer or the user to verify that the instrument is used in this environment.
Emissions test
Compliance
Electromagnetic environment – Guidelines
RF emissions CISPR 11
Group 1
Vision-R
TM
800 uses RF energy only for its internal
functions.
Therefore, its RF emissions are very low and are not
likely to generate any interference in nearby
electronic equipment.
RF emissions CISPR 11
Class B
Vision-R
TM
800 can be used in all sites, including
domestic sites and those that are directly connected
to the public low-voltage power supply network that
supplies private residential buildings.
Harmonic emissions
IEC 61000-3-2
Class A
Voltage fluctuations/Flicker
emissions IEC 61000-3-3
Complies
U
SER
MANUAL
> VIII. G
ENERAL
WARNINGS
87
Vision-R 800 - Essilor automatic phoropter > V1 - 12-2018
Summary of Contents for Vision-R 800
Page 1: ...www essilor instruments com User manual...
Page 4: ...I INTRODUCTION...
Page 8: ...II INSTRUMENT...
Page 17: ...III BASIC SETTINGS FOR PERFORMING AN EXAMINATION...
Page 36: ...IV PERFORMANCE OF SPECIFIC TESTS DURING A REFRACTION EXAMINATION...
Page 61: ...V INSTRUMENT SETTINGS...
Page 73: ...VI BREAKDOWN SERVICE...
Page 75: ...VII TECHNICAL DATA...
Page 78: ...VIII GENERAL WARNINGS...
Page 87: ...IX QR CODE...
Page 91: ...Essilor International 147 rue de Paris 94220 Charenton le Pont France www essilor com...