14
5.
After finishing work, take down the handle on the holder. When the temperature is less than 100
℃
, it enters
into the sleeping state.
CAUTION:
For prolong the life of the heater and protect the safety of the chip, use as low temperature as possible and as
big airflow as possible if it can finish the de-soldering work.
6.5 On Line with the Hot Plate
This unit can work together with a kind of the preheat plate by the special connection cord.
Take out the cord from the package. One end of the cord connects with the six-pins socket on the back of the unit,
and the other end connects with preheat plate. Detailed operation can refer to the manual of the preheat plate.
Ⅶ. Temperature Calibration
1.
If the unit is locked by password, it will not be able to calibrate the temperature and you must input the right
password.
2.
The corresponding temperature should be recalibrated after replacing the de-soldering gun or heating element
or nozzle or tip every time. It adopts digital calibration mode and input the revision value is input by pressing
button, make the adjustment simply and quickly.
3.
Method of recalibrating temperature: Use the thermometer to calibrate, and it is precise comparatively.
①
Set one kind tool’s temperature to a certain value(300
℃
).
②
When the temperature stabilizes, measure the temperature with thermometer and write down the reading.
③
In the single displaying state, press the “TOOL1” or “TOOL2” or “TOOL3” and not loosen, and then
press the
“
TEMP
▲”
and
“
TEMP
T
”
keys simultaneously, the unit enters into calibrating temperature
mode. The LCD shows “CAL” about 2seconds and then into the temperature-inputting interface.
z
Calibration the temperature of the soldering iron:
press the “TOOL1” key not loosen, and then
press the
“
TEMP
▲”
and
“
TEMP
T
”
keys simultaneously, the unit enters into the calibrating
temperature mode of the soldering iron.
z
Calibration the temperature of the de-soldering gun:
press the “TOOL2” key not loosen, and
then press the
“
TEMP
▲”
and
“
TEMP
T
”
keys simultaneously, the unit enters into the calibrating
temperature mode of the de-soldering gun.
z
Calibration the temperature of the SMD rework tool:
press the “TOOL3” key not loosen, and
then press the
“
TEMP
▲”
and
“
TEMP
T
”
keys simultaneously, the unit enters into the calibrating
temperature mode of the SMD rework tool.
④
At the moment, the digit of LCD display temperature is flashing. Press the
“
TEMP
▲”
or
“
TEMP
T
”
key to select the value and input the reading of thermometer, and the inputting method is the same as
temperature setting.
⑤
Press
“
VACUUM
”
key. Here, the whole calibration operation has been finished.
⑥
If the calibration is successful, the LCD shows “OK” and return to work state.
⑦
If the temperature still has deflection, you can repeat calibration in accordance with above steps.
z
Recommend using the 191/192 thermometer for measuring the temperature.