DESCRIPTION OF OPERATION
0740 800 186
- 19 -
© ESAB AB 2016
The processor on circuit board 20AP1 controls the HF generator and gas valve.
HF generator
When the welding torch switch is operated, and the open-circuit voltage is over 50 V, relay
RE2 closes and turns on the HF generator, TR2. It remains activated until the arc strikes, or
for a maximum of 0.5 seconds.
The voltage on the primary side of HF transformer 10TV1 is about 550 V. The secondary
voltage is about 11 kV if a 4 metre long welding torch is connected. If the welding torch is 16
metre long, the HF spark is about 8 kV.
Due to electromagnetic interference regulations, the energy in the HF ignition spark is
limited, and so the HF spark weakens with increasing length of the torch. The HF ignition is
satisfactory for welding torches up to 16 metres.
Gas valve
When the torch switch is operated, relay RE1 closes and energizes the gas valve.
When the torch switch is released and the gas post-flow time has elapsed, the gas valve is
deactivated.
TIG torch switch
The secondary windings of transformer TR1 and rectifier bridge D3-D6 produce 24 V DC.
This voltage energizes relay RE3 when the torch switch is closed.
10AP1 Component positions
WARNING!
Dangerous voltage - mains voltage.
Summary of Contents for Caddy Tig 1500i TA34
Page 10: ...WIRING DIAGRAM 0740 800 186 10 ESAB AB 2016 Tig 1500i with control panel TA34 ...
Page 11: ...WIRING DIAGRAM 0740 800 186 11 ESAB AB 2016 ...
Page 12: ...WIRING DIAGRAM 0740 800 186 12 ESAB AB 2016 Tig 2200i with control panels TA33 and TA34 ...
Page 13: ...WIRING DIAGRAM 0740 800 186 13 ESAB AB 2016 ...
Page 20: ...DESCRIPTION OF OPERATION 0740 800 186 20 ESAB AB 2016 ...