DESCRIPTION OF OPERATION
0740 800 186
- 16 -
© ESAB AB 2016
WARNING!
Dangerous voltage - mains voltage.
The primary circuit is supplied with 325 V DC from the power board, 15AP1. S1 is a fuse with
high rupturing capacity.
Transistor Q5 is the switching element in a switched voltage supply. The secondary voltage,
+24 V, is sensed by IC2 and controlled by IC1.
The isolation voltage of transformer T1 and optocoupler IC3 is 4 kV.
+15V A
Internal power supply to 2AP1 and power supply to the PFC circuits of the power board,
15AP1.
The voltage is controlled by VR1, it has a tolerance of ±0.75V.
2AP1:3 Secondary circuit
The secondary circuit delivers the following voltages:
+24 V
The voltage is controlled by IC1 on the primary side. It has a tolerance of ±0.6 V.
+5 V
The voltage is controlled by VR3. It has a tolerance of ±0.25 V.
-15 V
The voltage is controlled by VR2. It has a tolerance of ±0.75 V.