SpiroSphere
®
ECG Instructions for Use
Page 28/136
Version 02.02 • 04AUG2022
Preparing a Spirometry Measurement
Select a measurement
(e.g. <
Forced PRE Spirometry
>)
via tapping.
The measurement is started by tapping on <
Start
>.
The "
Ambient Conditions
" window appears:
Current room temperature (°C)
Current relative humidity (%)
Current barometric pressure (hPa)
Please observe the instructions for hygiene of your
system.
Verify that a new
ERT
PT with mouthpiece
is
attached in the SpiroSphere Sensor.
The current ambient conditions are to be entered manually. In this case, the
ambient data should be updated if the room temperature changed by more than
2°C or if relative humidity changed by more than 10%.
The ATP-BTPS correction factors for inspiratory and expiratory ows and
volumes will be determined from the ambient data. Therefore, ambient data
must be updated at regular intervals. Incorrect or imprecise ambient data will
result in incorrect measurement results.
Tap on <
Continue
> to apply the ambient data entered.
Make the proper preparations according to ATS/ERS guidance.