SpiroSphere
®
ECG Instructions for Use
Page 104/136
Version 02.02 • 04AUG2022
Malfunction
Cause of the Malfunction
Removal of the Malfunction
The ECG unit cannot be
switched on.
Battery discharged below
minimum power requirements.
Change battery for either a new or
(re)charged one.
Check polarity of the inserted
energy source.
The ECG unit is not
visible to the receiving
Bluetooth devices.
There is already a Bluetooth
pairing with another device.
Put the device into pairing mode.
For this, press the button after
power-on until the blue LED
ashes. This indicates that the
former pairing is removed and the
device is visible in the Bluetooth
vicinity.
No or fragmentary ECG
data is transferred.
Radio communication is
interrupted.
Reduce distance to receiving unit
or remove a possibly existing
obstacle.
Device detects no ECG
signal.
Check ECG cable application and
restart the device.
Remove and insert battery before
restart if normal restart leads to
the same malfunction.