SpiroSphere
®
ECG Instructions for Use
Page 123/136
Version 02.02 • 04AUG2022
Strong
electromagnetic elds can disturb the measurement signal.
Cellular phones, x-ray machines, thermal radio frequency treatment
equipment, diathermy, electrocautery, and RFID, etc. produce strong
electromagnetic elds. Make sure to keep devices that produce
electromagnetic elds away from the SpiroSphere device.
Make sure that the recommended distances to radio wave emitters are
met.
If the measurement signal is lost or degraded due to electromagnetic
disturbances, you need to relocate the device and repeat the
measurement.
Electromagnetic Interference and Wrong Diagnosis
The device must not be installed and used in the vicinity of high-
frequency devices, X-ray equipment, motors or transformers with high
installed power ratings since electric or magnetic interference elds may
falsify the measurement.
Do not perform measurements with the device directly next to other
equipment or in combination with other devices in a stacked form as this
may result in faulty operation.
The SpiroSphere has been tested for radiated RF immunity only at
selected frequencies. A use nearby emitters of other frequencies could
result in improper operation. Keep a distance of about 2 meters from
possible error sources when using the device.
Portable RF communications equipment (including peripherals such
as antenna cables and external antennas) should be used no closer
than 30 cm (12 inches) to any part of the medical device, including
cables speci ed by the manufacturer. Otherwise, degradation of the
performance of this equipment could result.
Use of accessories, cables etc. other than those speci ed or provided
by the manufacturer could result in increased electromagnetic emissions
or decreased electromagnetic immunity of this equipment and result in
improper operation.
Changes or modi cations to the SpiroSphere may result in increased
electromagnetic emissions or decreased electromagnetic immunity in
relation to EMC performance.
Devices that are connected to the SpiroSphere can also cause radiated
interference. Therefore, you must follow the regulatory and safety
related notes of the respective device manufacturer.
Summary of Contents for SpiroSphere
Page 1: ...781239 Version 02 02 Instructions for Use SpiroSphere ECG ...
Page 39: ...SpiroSphere ECG Instructions for Use Page 39 136 Version 02 02 04AUG2022 Pre Post Report ...
Page 134: ...SpiroSphere ECG Instructions for Use Page 134 136 Version 02 02 04AUG2022 For your notes ...
Page 135: ...SpiroSphere ECG Instructions for Use Page 135 136 Version 02 02 04AUG2022 ...