AFNOR NF X 15-211: 2009 standard
Filtration efficiency
Containment efficiency
Air face velocity
The classes established by the standard
Classification according to filtration type
The filtration efficiency
Class 1
Class 2
Normal operation
phase
Emissions concentration at the filter exhaust must be lower than 1%
of the TLV
Detection phase
The concentration at the filter exhaust must
be lower than 1% of the TLV, and the automatic
saturation detector must warn the user
The concentration at the filter exhaust
must be lower than 50% of the TLV
Safety operation phase
The concentration at the filter exhaust must be
lower than 50% of the TLV ; its duration must
not be lower than 1/12 of the normal operation
phase duration.
It is defined by the filter capacity to retain noxious molecules manipulated within
the enclosure and qualifies the quality of the recirculated air at the filter exhaust.
Class 1
Class 2
Filtration fume hood with safety reserve
Filtration fume hood without safety reserve
one main filtration level and
one safety filtration level
One filtration level
Denominations according to
the NF X 15 211:2009 standard
Denominations equivalence
for the erlab
®
products
Particles filtration*
Type P
Type P
Vapours filtration**
Type V
Type C
Particles and vapours filtration**
Type PV
Type PC
* : The particulate filter must be at least of H14 type according to the NF EN 1822-1 standard
** : The filters for vapours must be submitted to two successive performance tests with Cyclohexane and Isopropanol for the filters designed to retain
Volatile Organic Compounds (VOC). Another test for the acid vapours is carried out with Hydrochloric acid.
Appointed by the AFNOR, the Union de Normalisation de la Mécanique (UNM), composed of a college of
experts (INRS, national organizations, professional syndicates), has established the AFNOR NF X 15-211 :
2009 standard.
ETRAF) designed for research, analysis, education works,…. for all laboratories where chemicals subjected
to professional exposure limit values (OEL or TLV) are manipulated. This text imposes performances
criteria linked to:
Example of test carried out on a Captair
®
Flex™
XL 714, equipped with BE+ filters in class 1
Isopropanol
Cyclohexane
HCL (35%)
2250 gr
3204 gr
7862 gr
The retention
capacities recorded during the tests
performed on our filters demonstrate
the technical
performance developed by Erlab
®
.
These results ensure a very high protection level to the
Captair
®
Flex™ users.