User Manual
TM
12
Event 4:
Solvent detected 2
Event conditions:
The detection limit of the Molecode™ S unit has been surpassed.
CHANGE
FILTER(S)
Display screen
Alarm type
Solution :
It is necessary to replace the filter.
Please contact your maintenance service.
Solution :
Press the navigation and confirmation buttons simultaneously and release them to deacti-
vate the alarm.
Identify the source of the pollution.
Event 5:
Pollutant detected in the ambient air of the laboratory
Event conditions:
The ambient air sensor on the front detects that pollution is likely in the ambient air of
the laboratory.
LABORATORY
POLLUTION
Display screen
Alarm type
Red indicator light on, constant
Continuous audible alarm
Red indicator light on, blinking
Continuous audible alarm