7 Maintenance Information
44 |
44
201000401000 / Rev 04
7.3 Firmware Update
The secondary user performs the firmware update of the 1-channel transmitter. When software
faults have been identified and when new functions become available, it may be necessary to
update the firmware. The secondary user can obtain the firmware version of the 1-channel trans-
mitter using the maintenance software. Please contact your ergoline service partner for more
information.
7.4 Troubleshooting
FT1.4
Connect the 1-channel transmitter to the software.
Expected result: Once the 1-channel transmitter is connected, a charging icon
appears in the software. For charge levels above 40 % the icon is black.
FT1.5
Connect the 1-channel transmitter to the software. Cover the 1-channel trans-
mitter with your hand.
Expected result: Once the 1-channel transmitter is connected, a connection icon
appears in the software. The connection quality decreases. For levels below 40 %
the icon is red.
Error
Cause
1-channel transmitter cannot be
switched on
Check the battery charge level.
Check the contacts in the cover of the battery compart-
ment. If they are bent, replace the cover.
Check the battery clip. If it is bent, contact your ergoline
service partner.
Instable connection of the
1-channel transmitter
Check the battery charge level. A low battery charge level
may cause problems.
Check the position of the receiver.
Poor signal quality from 1-chan-
nel transmitter
Check the applied parts.
Defective applied parts may reduce the signal quality.
A wet chest belt may affect the signal quality.
Damp plugs of the adapter may affect the signal quality.
Damp plugs of the 1-channel transmitter may affect the
signal quality.