6 Cleaning and Disinfection Instructions
ers2 – 1-Channel ECG Telemetry System
39
201000401000 / Rev 04
6
Cleaning and Disinfection Instructions
Clean and disinfect the devices before their first use, before applying them to a patient, and after
removing them from a patient.
Clean and disinfect the devices when sending them to ergoline or a certified ergoline service partner.
Observe the following information for cleaning and disinfection.
6.1 1-Channel Transmitter
6.1.1
Cleaning
Recommended cleaning agent:
Isopropyl alcohol, 70 % solution (IPA)
• Prepare the cleaning solution according to the manufacturer's instructions.
• Dampen a clean cloth with the selected cleaning solution.
• Wring excess liquid from the cloth and squeeze until dry.
• Wipe the exterior surfaces of the 1-channel transmitter until all visible soil has been removed.
NOTICE
Inappropriate cleaning and disinfection
When the battery compartment of the 1-channel transmitter is not properly closed, humidity
entering the compartment during cleaning and disinfection may damage the device.
Close the battery compartment of the 1-channel transmitter before cleaning and disinfection
according to the instructions in this manual.
The protection rating applies only if the following conditions are met