![ergoline ers2 System Manual Download Page 42](http://html.mh-extra.com/html/ergoline/ers2/ers2_system-manual_2422185042.webp)
6 Cleaning and Disinfection Instructions
40 |
40
201000401000 / Rev 04
6.1.2
Disinfection
Clean the 1-channel transmitter before disinfecting it.
Recommended disinfectant:
Isopropyl alcohol (IPA), 70 % solution, spray
• Apply IPA spray to the surface from a distance of 15 to 20 cm until the entire surface is wet.
• Allow the solution to remain on the treated surface for at least 1 minute.
• Dry the treated surface with a clean cloth.
• Dampen a clean cloth with distilled water.
• Wring excess liquid from the cloth.
• Wipe the exterior surfaces of the 1-channel transmitter.
• Use another clean cloth to dry the surface of the 1-channel transmitter.
6.2 Adapter
6.2.1
Cleaning
Recommended cleaning agent:
Isopropyl alcohol, 70 % solution (IPA)
• Prepare the cleaning solution according to the manufacturer's instructions.
• Dampen a clean cloth with the selected cleaning solution.
• Wring excess liquid from the cloth and squeeze until dry.
• Wipe the exterior surfaces of the adapter until all visible soil has been removed.
6.2.2
Disinfection
Clean the adapter before disinfecting it.
Recommended disinfectant:
Isopropyl alcohol, 70 % solution (IPA spray)
• Apply IPA spray to the surface from a distance of 15 to 20 cm until the entire surface is wet.
• Allow the solution to remain on the treated surface for at least 1 minute.
• Dry the treated surface with a clean cloth.
• Repeat steps 1 to 3.
• Dampen a clean cloth with distilled water.
• Wring excess liquid from the cloth.
• Wipe the exterior surfaces of the adapter.