5 Attaching the ECG System to the Patient
32 |
32
201000401000 / Rev 04
5
Attaching the ECG System to the Patient
5.1 Visual Inspection of the ECG System
Visually inspect the ECG system before each use. If one of the checks fails, please contact your
ergoline service partner for further information.
CAUTION
Incorrect application
The ECG system is not intended for long-term monitoring. Skin irritation and/or allergies may occur.
Apply the ECG system only as described in this manual.
CAUTION
Insufficient ECG transmission quality
Line isolation monitor transients may resemble actual cardiac waveforms and thus inhibit heart
rate alarms. This problem can be largely avoided by correct application of the chest belt and/or
adapter.
Regularly check the 1-channel transmitter for correct application to the chest belt or adapter and
patient according to the instructions in this manual.
CAUTION
Risk of infection
After use, the ECG system may be contaminated with bacteria or viruses.
Clean and disinfect the 1-channel transmitter and the applied parts according to the cleaning and
disinfection instructions in this manual.
CAUTION
Risk of infection
The ECG system may only be applied on intact skin (e.g., not over open wounds, lesions, infected or
inflamed areas).
CAUTION
Wet environment
There is a risk of burn injuries and insufficient ECG transmission quality.
The ECG system is not suitable for use in highly wet environments. Do not wear it in the shower or
when swimming.
WARNING
Visual inspection before use
Before each use, visually inspect the ECG system for signs of damage. If you detect damage that may
result in a hazard to the patient or the operator, the ECG system must be repaired before it can be
used again. Follow the instructions for visual inspection given in this manual.
Visual Inspection
Pass
Fail
Check the housing for mechanical damage, such as cracks, scuffing or
distortion due to heat.
Expected result: no damage
Shake the device and listen for any loose or detached components.
Expected result: no sound
Check the device and snap fastener contacts for contamination, dust,
and humidity. Expected result: no contamination, no humidity
Check cover of battery compartment and sealing ring for wear.
Expected result: no bent contacts on the inside, sealing ring flush on
housing