23
5.9 Cleaning and disinfection by machine
The washer/disinfector must have fundamentally certified ef-
fectiveness (e.g. CE marking according to DIN EN ISO 15883).
The detergent and disinfectant must be suitable for medical de-
vices made of plastic and metal and have a pH value between 5.5
and 11.
It is essential to follow the cleaning agent and disinfectant man-
ufacturers’ instructions.
1. Roll up the cable loosely and lay the product carefully in a
suitable rinsing basket. In doing so, make sure the product is
not touching other instruments or parts of instruments.
2. Start a tested program with the following properties:
– Thermal disinfection: 5 to 10 minutes at 90 to 93 °C,
A
0
≥
3000.
– Final rinse with distilled or deionized water.
– Sufficient product drying.
3. If contaminants are still visible on the product after the end
of the routine, repeat precleaning and machine cleaning and
disinfection.
5.10 Check
1. Check the product for visible damage and wear:
– Damage to the product, e.g. cracks, rough surface, splin-
tering.
– Damage to the product insulation and/or cable/connec-
tor insulation, e.g. cracks and breaks.
2. Check the mechanical functionality of the product:
– Can all buttons be pressed?
If damaged, do not use this product!
OBJ_DOKU-182711-004.fm Seite 23 Dienstag, 22. November 2016 3:25 15
Summary of Contents for Nervtest 20190-148
Page 2: ...OBJ_DOKU 182711 004 fm Seite 2 Dienstag 22 November 2016 3 25 15...
Page 14: ...14 OBJ_DOKU 182711 004 fm Seite 14 Dienstag 22 November 2016 3 25 15...
Page 36: ...36 OBJ_DOKU 182711 004 fm Seite 36 Dienstag 22 November 2016 3 25 15...
Page 48: ...48 OBJ_DOKU 182711 004 fm Seite 48 Dienstag 22 November 2016 3 25 15...
Page 60: ...60 OBJ_DOKU 182711 004 fm Seite 60 Dienstag 22 November 2016 3 25 15...
Page 72: ...72 OBJ_DOKU 182711 004 fm Seite 72 Dienstag 22 November 2016 3 25 15...
Page 73: ...EL 20190 148 OBJ_DOKU 182711 004 fm Seite 73 Dienstag 22 November 2016 3 25 15...
Page 76: ...76 3 Erbe Elektromedizin OBJ_DOKU 182711 004 fm Seite 76 Dienstag 22 November 2016 3 25 15...
Page 80: ...80 5 5 1 5 6 5 7 1 5 8 2 bar OBJ_DOKU 182711 004 fm Seite 80 Dienstag 22 November 2016 3 25 15...
Page 84: ...84 OBJ_DOKU 182711 004 fm Seite 84 Dienstag 22 November 2016 3 25 15...
Page 96: ...96 OBJ_DOKU 182711 004 fm Seite 96 Dienstag 22 November 2016 3 25 15...
Page 108: ...108 OBJ_DOKU 182711 004 fm Seite 108 Dienstag 22 November 2016 3 25 15...
Page 120: ...120 OBJ_DOKU 182711 004 fm Seite 120 Dienstag 22 November 2016 3 25 15...
Page 142: ...142 OBJ_DOKU 182711 004 fm Seite 142 Dienstag 22 November 2016 3 25 15...
Page 164: ...164 OBJ_DOKU 182711 004 fm Seite 164 Dienstag 22 November 2016 3 25 15...
Page 165: ...RU 20190 148 OBJ_DOKU 182711 004 fm Seite 165 Dienstag 22 November 2016 3 25 15...
Page 168: ...168 3 Erbe Elektromedizin OBJ_DOKU 182711 004 fm Seite 168 Dienstag 22 November 2016 3 25 15...
Page 172: ...172 5 5 1 5 6 5 7 1 2 OBJ_DOKU 182711 004 fm Seite 172 Dienstag 22 November 2016 3 25 15...
Page 176: ...176 OBJ_DOKU 182711 004 fm Seite 176 Dienstag 22 November 2016 3 25 15...
Page 187: ...ZH 20190 148 OBJ_DOKU 182711 004 fm Seite 187 Dienstag 22 November 2016 3 25 15...
Page 196: ...196 OBJ_DOKU 182711 004 fm Seite 196 Dienstag 22 November 2016 3 25 15...
Page 197: ...OBJ_DOKU 182711 004 fm Seite 197 Dienstag 22 November 2016 3 25 15...