17

IMPORTANT!

Please read all information carefully.
These instructions for use do not replace the user manual of the 

electrosurgical unit used! Read the user manual of the electro-
surgical unit and ask Erbe or your distributor in case of doubt!

Normal use

The electrode handle Nervtest is intended for the localization of 

electrically stimulable tissue structures, e.g. facial nerve.
The electrode handle Nervtest may only be used in combination 
with the silver ball electrodes (art. no. 20191-079 or 21191-
052).
The electrode handle Nervtest is intended to be connected to the 

following Erbe units only:

ICC 350 ZMK in conjunction with remote control FB2 Art.-Nr. 
20121-007 and an HF neutral electrode

NT 2 in conjunction with the NT2 disposable return electrode 
(Art. No. 20142-200) and the NT2 neutral electrode cable 

(Art. No. 20142-100, 20142-101)

Maximum electrical capacity

The maximum electrical capacity is given on the instrument in 
[V

p

].

OBJ_DOKU-182711-004.fm  Seite 17  Dienstag, 22. November 2016  3:25 15

Summary of Contents for Nervtest 20190-148

Page 1: ...1 73 Elektrodengreep zenuwtest 85 Elektrodeh ndtag Nervetest 97 Elektrodhandtag f r nervtest 109 Hermotesti elektrodikahva 121 Uchwyt elektrody neurotest 131 Dr k elektrody Nervtest 143 Nervtest elekt...

Page 2: ...OBJ_DOKU 182711 004 fm Seite 2 Dienstag 22 November 2016 3 25 15...

Page 3: ...DE VERWENDUNGSHINWEIS Elektrodengriff Nervtest 20190 148 OBJ_DOKU 182711 004 fm Seite 3 Dienstag 22 November 2016 3 25 15...

Page 4: ...odukt anschlie en 7 5 Reinigung Desinfektion Sterilisation 8 5 1 Sicherheitshinweise 8 5 2 Begrenzung der Wiederaufbereitung 8 5 3 Empfohlene Ausstattung Mittel 9 5 4 Empfohlene Verfahren 9 5 5 Zerleg...

Page 5: ...trodengriff Nervtest darf nur in Kombination mit den Sil berkugelelektroden Art Nr 20191 079 oder 21191 052 ver wendet werden Der Elektrodengriff Nervtest ist ausschlie lich zum Anschluss an folgende...

Page 6: ...chanische Sch den aufweist Erbe Elektromedizin empfiehlt eine elektrische Durchgangspr fung vor jeder Anwendung dieses Produkts durchzuf hren Hinweis Falls der Kabelleiter gebrochen ist oder der elekt...

Page 7: ...ndung die Isolierung der An schlusskabel die Aktivierungsfunktionen des Produkts und ob das Produkt mechanische Sch den aufweist Falls schadhaft dieses Produkt nicht verwenden 4 3 Produkt montieren St...

Page 8: ...gesp lt wer den Maximaler Luftdruck beim Trocknen 2 bar Maximaler Wasserdruck beim Sp len 2 bar Bei der maschinellen Reinigung Desinfektion 95 C nicht ber schreiten Angaben des Reinigungs und Desinfek...

Page 9: ...tor G 7836 CD Miele Cie KG G tersloh Deutschland mit dem Reinigungsmittel neodisher mediclean forte Dr Weigert GmbH Co KG Hamburg Deutschland gem empfohlenem Waschprogramm 5 4 Empfohlene Verfahren Erb...

Page 10: ...eignet ist Verwenden Sie ein zum Reinigungsmittel kompatibles Desinfek tionsmittel das f r die Bereitung eines Tauchbads geeignet ist Das Reinigungs und das Desinfektionsmittel m ssen f r Medi zinprod...

Page 11: ...Halten Sie die empfohlene Einwirkzeit ein 7 Sp len Sie die Oberfl chen mit sterilem deionisierten Wasser mind 1 Minute 8 Trocknen Sie das Produkt mit gefilterter Druckluft 5 9 Maschinelle Reinigung un...

Page 12: ...n Sie das Produkt auf sichtbare Besch digungen und Verschlei Besch digungen am Produkt z B Risse raue Oberfl che Absplitterungen Besch digungen an der Isolation des Produkts und oder des Kabels Stecke...

Page 13: ...lektromedizin Dampfsterilisation Fraktioniertes Vakuumverfahren mit ausreichender Produkt trocknung Haltezeit 3 bis 18 Minuten bei 132 bis 135 C Sterilisator gem geltender nationaler Normen und Vor sc...

Page 14: ...14 OBJ_DOKU 182711 004 fm Seite 14 Dienstag 22 November 2016 3 25 15...

Page 15: ...EN NOTES ON USE Electrode handle Nervtest 20190 148 OBJ_DOKU 182711 004 fm Seite 15 Dienstag 22 November 2016 3 25 15...

Page 16: ...up the product 19 5 Cleaning disinfection sterilization 20 5 1 Safety instructions 20 5 2 Reconditioning limitation 20 5 3 Recommended equipment substances 20 5 4 Recommended methods 21 5 5 Dismantli...

Page 17: ...ndle Nervtest may only be used in combination with the silver ball electrodes art no 20191 079 or 21191 052 The electrode handle Nervtest is intended to be connected to the following Erbe units only I...

Page 18: ...ectrical con tinuity test before every use of this product Note If the cable conductor is broken or the electrical continuity of the cable is interrupted in any other manner it can cause arc ing in th...

Page 19: ...nsert the silver ball electrode art no 20191 079 or 21191 052 into the electrode handle Nervtest art no 20190 148 4 4 Connect up the product Connection to an ICC 350 ZMK Insert the handle into the Neu...

Page 20: ...ning can damage this product Observe the safety instructions for checking the product With visible damage or functional impairment the product may no longer be used 5 3 Recommended equipment substance...

Page 21: ...ne monitoring check of material compatibility 5 5 Dismantling 1 Remove the HF electrode from the handle 5 6 Required Aids 5 7 Precleaning Use water for precleaning a non fixing disinfectant if necessa...

Page 22: ...instructions 2 Immerse the product completely in the cleaning bath with out the product touching other parts in the bath Observe the recommended immersion time 3 Thoroughly clean the surfaces with a...

Page 23: ...s or parts of instruments 2 Start a tested program with the following properties Thermal disinfection 5 to 10 minutes at 90 to 93 C A0 3000 Final rinse with distilled or deionized water Sufficient pro...

Page 24: ...zation methods are used Erbe Elektromedizin shall bear no responsibility Steam sterilization Fractionated vacuum method with adequate product drying Holding time 3 to 18 minutes at 132 to 135 C Steril...

Page 25: ...FR NOTICE D UTILISATION Manche de test nerveux 20190 148 OBJ_DOKU 182711 004 fm Seite 25 Dienstag 22 November 2016 3 25 15...

Page 26: ...it 29 4 4 Connecter le produit 29 5 Nettoyage d sinfection et st rilisation 30 5 1 Consignes de s curit 30 5 2 Limitation de la pr paration 30 5 3 Equipement Moyens recommand s 31 5 4 Proc d s recomma...

Page 27: ...exemple le nerf facial Le manche de test nerveux doit tre utilis uniquement avec les lectrodes boules argent es r f 20191 079 ou 21191 052 Le manche de test nerveux est destin tre connect unique ment...

Page 28: ...ensemble Erbe recommande de proc der un contr le de continuit lec trique avant chaque utilisation de ce produit Remarque si le conducteur du c ble est bris ou la continuit lectrique interrompue pour t...

Page 29: ...ant chaque emploi l isolation des c bles de raccorde ment les fonctions de commande et le parfait tat m canique de l ensemble Ne pas utiliser ce produit s il est endommag 4 3 Monter le produit Embo te...

Page 30: ...e s chage 2 bars Pression maximale de l eau pour le lavage 2 bars La temp rature de nettoyage d sinfection en machine ne doit pas d passer 95 C Respecter imp rativement les instructions du fabricant d...

Page 31: ...c un d sinfecteur G 7836 CD Miele Cie KG G tersloh Allemagne et le d tergent neodisher mediclean forte Dr Weigert GmbH Co KG Hambourg Allemagne conform ment au pro gramme de lavage recommand 5 4 Proc...

Page 32: ...fectant compatible avec le produit de nettoyage et convenant la pr paration d un bain de trempage Le produit de nettoyage et de d sinfection doit convenir pour les dispositifs m dicaux en mati re plas...

Page 33: ...se trouvant dans le bain Laisser tremper le produit pendant le temps recom mand 7 Rincer les surfaces l eau d sionis e st rile pendant au mi nimum 1 minute 8 S cher le produit l air comprim filtr 5 9...

Page 34: ...ges et d usure visibles sur le produit Dommages sur le produit par exemple fissures surfaces rugueuses caillements Dommages sur l isolation du produit et ou du c ble du raccord par exemple fissures et...

Page 35: ...tion St rilisation la vapeur Vide fractionn avec s chage suffisant du produit Dur e 3 18 minutes 132 135 C St rilisateur conforme aux normes et r glements nationaux en vigueur p ex DIN EN 13060 ou DIN...

Page 36: ...36 OBJ_DOKU 182711 004 fm Seite 36 Dienstag 22 November 2016 3 25 15...

Page 37: ...ES INDICACI N DE USO Mango de Nervtest 20190 148 OBJ_DOKU 182711 004 fm Seite 37 Dienstag 22 November 2016 3 25 15...

Page 38: ...ar el producto 41 5 Limpieza desinfecci n esterilizaci n 42 5 1 Indicaciones de seguridad 42 5 2 Limitaci n del reacondicionamiento 42 5 3 Equipos y productos recomendados 43 5 4 Procedimientos recome...

Page 39: ...ervtest solo se debe utilizaren combinaci n con los electrodos de bola de plata n art 20191 079 o 21191 052 El mango de electrodo para Nervtest est previsto nicamente para su conexi n a los equipos Er...

Page 40: ...romedizin recomienda realizar un control de continui dad el ctrica antes de cada uso de este producto Indicaci n En caso de que el conductor est deteriorado o el paso de corriente el ctrica est interr...

Page 41: ...obar el producto Verifique Ud antes de cada utilizaci n el aislamiento de los ca bles de conexi n las funciones de activaci n y la posible existen cia de deterioros mec nicos Si este producto est dete...

Page 42: ...uso Presi n m xima del aire durante el secado 2 bar Presi n m xima del agua durante el lavado 2 bar Durante la limpieza desinfecci n mec nica la temperatura no debe superar los 95 C Observe estrictam...

Page 43: ...ci n G 7836 CD Miele Cie KG G tersloh Alemania con el producto de limpieza neodisher mediclean forte Dr Weigert GmbH Co KG Hamburg Alemania seg n el programa de lavado recomendado 5 4 Procedimientos r...

Page 44: ...ante compatible con el producto de limpieza y apto para ba o de inmersi n Los productos de limpieza y de desinfecci n deben ser adecua dos para productos sanitarios de pl stico o metal y tener un pH e...

Page 45: ...a o Observe el tiempo de actuaci n recomendado 7 Lave las superficies con agua desionizada est ril como m n durante 1 minuto 8 Seque el producto con aire comprimido filtrado 5 9 Limpieza mec nica y de...

Page 46: ...be que el producto no presente da os ni desgastes visibles Deterioros en el producto p ej fisuras superficies ru gosas astillamientos Deterioros en el aislamiento del producto y o del cable enchufe p...

Page 47: ...i n por vapor Procedimiento de vac o fraccionado con secado suficiente del producto Tiempo de permanencia de 3 a 18 minutos a una tempera tura de 132 a 135 C Esterilizador seg n las normas y disposici...

Page 48: ...48 OBJ_DOKU 182711 004 fm Seite 48 Dienstag 22 November 2016 3 25 15...

Page 49: ...IT INDICAZIONI PER L IMPIEGO Impugnatura per elettrodi Nervtest 20190 148 OBJ_DOKU 182711 004 fm Seite 49 Dienstag 22 November 2016 3 25 15...

Page 50: ...legamento del prodotto 53 5 Pulizia disinfezione e sterilizzazione 54 5 1 Indicazioni di sicurezza 54 5 2 Limiti del ricondizionamento 54 5 3 Dotazione strumenti raccomandati 55 5 4 Metodo raccomandat...

Page 51: ...o faccia le L impugnatura per elettrodi Nervtest pu essere utilizzata esclu sivamente in abbinamento agli elettrodi a sfere in argento art n 20191 079 o 21191 052 L impugnatura per elettrodi Nervtest...

Page 52: ...uire un test elettrico prima di ogni utilizzo di questo prodotto Nota se il conduttore del cavo rotto o se la continuit elettrica del cavo interrotta in altro modo nella linea di ritorno dal pa ziente...

Page 53: ...zzare il prodotto se danneggiato 4 3 Montaggio del prodotto Inserire l elettrodo a sfera in argento art n 20191 079 o 21191 052 nell impugnatura Nervtest art n 20190 148 4 4 Collegamento del prodotto...

Page 54: ...tura 2 bar Pressione massima dell acqua durante il risciacquo 2 bar Non superare i 95 C durante la pulizia la disinfezione meccani ca Attenersi scrupolosamente alle indicazioni del produttore del dete...

Page 55: ...n apparecchio G 7836 CD Miele Cie KG G tersloh Germania con il detergente neodisher mediclean forte Dr Weigert GmbH Co KG Amburgo Germania secondo il programma di lavaggio con sigliato 5 4 Metodo racc...

Page 56: ...e Utilizzare un disinfettante compatibile con il detergente adatto all impiego nel bagno a immersione I detergenti e i disinfettanti devono essere adatti a dispositivi medici in plastica e metallo e p...

Page 57: ...nenti Attenersi al tempo di esposizione raccomanda to 7 Sciacquare le superfici min 1 minuto con acqua sterile deionizzata 8 Asciugare il prodotto utilizzando aria compressa filtrata 5 9 Pulizia e dis...

Page 58: ...ione meccanica 5 10 Controllo 1 Controllare che il prodotto non presenti danni e usura visibili Danni visibili sul prodotto ad es crepe superfici ruvide scheggiature Danni all isolamento del prodotto...

Page 59: ...ore Procedimento in vuoto frazionato con sufficiente asciugatura del prodotto Tempo di permanenza da 3 a 18 minuti a una temperatura compresa tra 132 e 135 C Sterilizzatore conforme alle norme e dispo...

Page 60: ...60 OBJ_DOKU 182711 004 fm Seite 60 Dienstag 22 November 2016 3 25 15...

Page 61: ...PT INDICA O DE UTILIZA O Punho de el ctrodo de Nervtest 20190 148 OBJ_DOKU 182711 004 fm Seite 61 Dienstag 22 November 2016 3 25 15...

Page 62: ...r o produto 65 5 Limpeza desinfec o esteriliza o 66 5 1 Indica es de seguran a 66 5 2 Limites para a reutiliza o 66 5 3 Equipamentos meios recomendados 67 5 4 M todos recomendados 67 5 5 Desmontagem 6...

Page 63: ...vo facial O punho de el trodo de Nervtest s pode ser usado em combi na o com o el trodo esf rico de prata art n 20191 079 ou 21191 052 O punho de el ctrodo Nervtest destina se exclusivamente li ga o a...

Page 64: ...medizin recomenda submeter o produto a um teste de continuidade el ctrica antes de cada utiliza o Nota No caso de o condutor do cabo estar quebrado ou a pas sagem el ctrica do mesmo estar de algum mod...

Page 65: ...trolar o produto Antes de cada utiliza o controle o isolamento dos cabos de li ga o as fun es de activa o e eventuais danifica es mec nicas no produto N o utilizar este produto caso este apresente def...

Page 66: ...o atmosf rica m xima durante a secagem 2 bar Press o de gua m xima para a lavagem 2 bar Durante a limpeza desinfec o autom tica n o ultrapassar a temperatura de 95 C Respeite impreterivelmente as ins...

Page 67: ...desinfector G 7836 CD Miele Cie KG G tersloh Alemanha com o produto de limpeza neodisher mediclean forte Dr Weigert GmbH Co KG Hamburgo Alemanha de acordo com o programa de lavagem recomendado 5 4 M t...

Page 68: ...c o compat vel com o produto de limpeza e que seja adequado para a prepara o de um banho de imers o Os produtos de limpeza e desinfec o devem ser adequados para dispositivos m dicos de pl stico e meta...

Page 69: ...es no banho Cumpra o tempo de actua o recomendado 7 Lave as superf cies com gua desionizada est ril pelo me nos 1 minuto 8 Seque o produto com ar comprimido filtrado 5 9 Limpeza e desinfec o mec nicas...

Page 70: ...e se o produto apresenta quaisquer danos vis veis e ou pontos de desgaste Danos no produto p ex fissuras superf cie rugosa las cada Danos no isolamento do produto e ou do cabo da ficha p ex fissuras e...

Page 71: ...vapor M todo de v cuo fraccionado com secagem suficiente do produto Tempo de perman ncia de 3 a 18 minutos entre 132 e 135 C Esterilizador de acordo com as normas e regulamentos na cionais em vigor p...

Page 72: ...72 OBJ_DOKU 182711 004 fm Seite 72 Dienstag 22 November 2016 3 25 15...

Page 73: ...EL 20190 148 OBJ_DOKU 182711 004 fm Seite 73 Dienstag 22 November 2016 3 25 15...

Page 74: ...1 75 2 75 3 76 4 77 4 1 77 4 2 77 4 3 77 4 4 78 5 78 5 1 78 5 2 79 5 3 79 5 4 79 5 5 80 5 6 80 5 7 80 5 8 80 5 9 81 5 10 82 5 11 82 5 12 83 OBJ_DOKU 182711 004 fm Seite 74 Dienstag 22 November 2016 3...

Page 75: ...75 Erbe 1 20191 079 21191 052 Erbe ICC 350 ZMK FB2 20121 007 HF NT 2 NT2 20142 200 NT2 20142 100 20142 101 2 Vp OBJ_DOKU 182711 004 fm Seite 75 Dienstag 22 November 2016 3 25 15...

Page 76: ...76 3 Erbe Elektromedizin OBJ_DOKU 182711 004 fm Seite 76 Dienstag 22 November 2016 3 25 15...

Page 77: ...77 RBE Elektromedizin RBE Elektromedizin 4 4 1 4 2 4 3 20191 079 21191 052 20190 148 OBJ_DOKU 182711 004 fm Seite 77 Dienstag 22 November 2016 3 25 15...

Page 78: ...78 4 4 ICC 350 ZMK Erbe ICC 350 ZMK Erbe ICC 350 ZMK 1 5 NT 2 NT 2 5 5 1 2 bar 2 bar 95 C OBJ_DOKU 182711 004 fm Seite 78 Dienstag 22 November 2016 3 25 15...

Page 79: ...Cidex OPA Johnson Johnson Medical Limited Gargrave Skipton M G 7836 CD Miele Cie KG Guetersloh neodisher mediclean forte Dr Weigert GmbH Co KG Hamburg 5 4 Erbe OBJ_DOKU 182711 004 fm Seite 79 Dienstag...

Page 80: ...80 5 5 1 5 6 5 7 1 5 8 2 bar OBJ_DOKU 182711 004 fm Seite 80 Dienstag 22 November 2016 3 25 15...

Page 81: ...81 pH 5 5 11 1 2 3 4 1 5 6 7 1 8 5 9 CE DIN EN ISO 15883 pH 5 5 11 OBJ_DOKU 182711 004 fm Seite 81 Dienstag 22 November 2016 3 25 15...

Page 82: ...82 1 2 5 10 90 93 C A0 3000 3 5 10 1 2 5 11 1 OBJ_DOKU 182711 004 fm Seite 82 Dienstag 22 November 2016 3 25 15...

Page 83: ...83 5 12 Erbe Elektromedizin Erbe Elektromedizin 3 18 132 135 C DIN EN 13060 DIN EN 285 DIN EN ISO 17665 OBJ_DOKU 182711 004 fm Seite 83 Dienstag 22 November 2016 3 25 15...

Page 84: ...84 OBJ_DOKU 182711 004 fm Seite 84 Dienstag 22 November 2016 3 25 15...

Page 85: ...NL TOEPASSINGSAANWIJZING Elektrodengreep zenuwtest 20190 148 OBJ_DOKU 182711 004 fm Seite 85 Dienstag 22 November 2016 3 25 15...

Page 86: ...4 4 Product aansluiten 89 5 Reiniging desinfectie sterilisatie 90 5 1 Veiligheidsinstructies 90 5 2 Beperking van gereed maken voor gebruik 90 5 3 Aanbevolen verzorging middelen 91 5 4 Aanbevolen proc...

Page 87: ...greep zenuwtest mag uitsluitend in combinatie met de zilveren kogelelektrode art nr 20191 079 of 21191 052 worden gebruikt De elektrodengreep zenuwtest is uitsluitend bestemd voor aan sluiting op de v...

Page 88: ...oont Erbe Elektromedizin adviseert v r ieder gebruik van dit product een elektrische doorgangsmeting uit te voeren Aanwijzing Indien de kabelgeleider is gebroken of de elektrische doorgang van de kabe...

Page 89: ...e activeringsfuncties van het product en of het product mecha nische beschadigingen vertoont Bij beschadigingen dit product niet gebruiken 4 3 Product monteren Steek de zilveren kogelelektrode art nr...

Page 90: ...imale waterdruk tijdens het spoelen 2 bar Bij machinale reiniging desinfectie 95 C niet overschrijden De voorschriften van de producent van het desinfecteermiddel beslist in acht nemen Bij sterilisati...

Page 91: ...6 CD Miele Cie KG G tersloh Duitsland met het reinigingsmiddel neodisher mediclean forte Dr Weigert GmbH Co KG Hamburg Duitsland conform het aanbevo len wasprogramma 5 4 Aanbevolen procedures Erbe adv...

Page 92: ...ebruik een desinfecterend middel dat compatibel is met het reinigingsmiddel en dat geschikt is voor het bereiden van een dompelbad Reinigings en desinfecterend middel moeten geschikt zijn voor medisch...

Page 93: ...met andere onderdelen in het bad Houd u zich aan de aanbevolen inwerktijd 7 Spoel de oppervlakken met steriel gede oniseerd water min 1 minuut 8 Droog het product met gefilterde perslucht 5 9 Machinal...

Page 94: ...ntroleer het product op zichtbare beschadigingen en slij tage Beschadigingen aan het product bijv scheuren ruwe oppervlakken splinters Beschadigingen aan de isolatie van het product en of de kabel ste...

Page 95: ...van Erbe Elektromedizin Stoomsterilisatie Gefractioneerd vacu mproced met voldoende productdro ging Verblijftijd 3 tot 18 minuten bij 132 tot 135 C Sterilisator volgens geldende nationale normen en vo...

Page 96: ...96 OBJ_DOKU 182711 004 fm Seite 96 Dienstag 22 November 2016 3 25 15...

Page 97: ...DA ANVENDELSESHENVISNING Elektrodeh ndtag Nervetest 20190 148 OBJ_DOKU 182711 004 fm Seite 97 Dienstag 22 November 2016 3 25 15...

Page 98: ...Tilslutning af produktet 101 5 Reng ring desinfektion sterilisation 102 5 1 Sikkerhedsanvisninger 102 5 2 Begr nsning for ny klarg ring 102 5 3 Anbefalet udstyr midler 103 5 4 Anbefalet procedure 103...

Page 99: ...trodeh ndtaget Nervetest m kun anvendes i kombination med s lvkugleelektroder art nr 20191 079 eller 21191 052 Elektrodeh ndtaget Nervetest er udelukkende beregnet til til slutning til f lgende Erbe a...

Page 100: ...er Erbe Elektromedizin anbefaler at der foretages en elektrisk gen nemgangstest f r hver anvendelse af dette produkt Bem rk Hvis kabellederen er kn kket eller hvis kablets elektri ske forbindelse p an...

Page 101: ...gskablerne produktets aktiveringsfunktioner samt om der forekommer mekaniske beskadigelser Produktet m ikke anvendes hvis det er beskadiget 4 3 Montering af produktet Anbring s lvkugleelektroden art n...

Page 102: ...ing 2 bar Maksimalt vandtryk ved skylning 2 bar Ved den maskinelle reng ring desinfektion m 95 C ikke over skrides Reng rings og desinfektionsmiddelproducenternes angivelser skal ubetinget overholdes...

Page 103: ...836 CD Miele Cie KG G tersloh Tyksland med reng ringsmidlet neodisher mediclean forte Dr Weigert GmbH Co KG Hamburg Tyksland i overensstemmelse med anbefalet va skeprogram 5 4 Anbefalet procedure Erbe...

Page 104: ...ingsbad Der b r anvendes et desinfektionsmiddel kompatibelt med ren g ringsmidlet der er velegnet til tilberedelsen af et neddyp ningsbad Reng rings og desinfektionsmidlet skal v re egnet til medicin...

Page 105: ...e handlingstid skal overholdes 7 Skyl overfladerne med sterilt deioniseret vand mindst 1 minut 8 Produktet t rres ved hj lp af filtreret trykluft 5 9 Maskinel reng ring og desinfektion Reng rings og d...

Page 106: ...10 Kontrol 1 Produktet kontrolleres for synlige beskadigelser og slitage Beskadigelser p produktet fx ridser ru overflade afsl ede kanter Beskadigelser p isoleringen af produktet og eller ka bel stik...

Page 107: ...n Fraktioneret vakuumprocedure med tilstr kkelig produkt t rring Holdetid 3 til 18 minutter ved 132 til 135 C Sterilisator i henhold til g ldende nationale standarder og forskrifter f eks DIN EN 13060...

Page 108: ...108 OBJ_DOKU 182711 004 fm Seite 108 Dienstag 22 November 2016 3 25 15...

Page 109: ...SV ANV NDARHANDLEDNING Elektrodhandtag f r nervtest 20190 148 OBJ_DOKU 182711 004 fm Seite 109 Dienstag 22 November 2016 3 25 15...

Page 110: ...kten 113 5 Reng ring desinfektion sterilisering 114 5 1 S kerhetsanvisningar 114 5 2 Begr nsningar i teranv ndande 114 5 3 Rekommenderad utrustning rekommenderade medel 115 5 4 Rekommendationer 115 5...

Page 111: ...trodhandtaget f r nervtest f r endast anv ndas i kombina tion med silverkuleelektroden art nr 20191 079 eller 21191 052 Elektrodhandtaget f r nervtest r endast avsett att anslutas till f ljande Erbe u...

Page 112: ...elektrisk lednings kontroll utf rs f re varje anv ndning Information Om kabelledaren r avbruten eller om kabelns elek triska passage har avbrutits p n got annat s tt kan det bildas en ljusb ge i pati...

Page 113: ...g produktens aktive ringsfunktioner och om produkten har mekaniska skador f re varje anv ndning Anv nd inte produkten om den r skadad 4 3 Montering av produkten F r in silverkuleelektroden art nr 2019...

Page 114: ...r Maximalt vattentryck vid spolning 2 bar Vid maskinell reng ring desinficering f r 95 C inte verskridas Uppgifterna fr n tillverkaren av reng rings och desinfektions medlen skall ovillkorligen beakta...

Page 115: ...ektor G 7836 CD Miele Cie KG G tersloh Tyskland med ren g ringsmedlet neodisher mediclean forte Dr Weigert GmbH Co KG Hamburg Tyskland enligt rekommenderat tv ttprogram 5 4 Rekommendationer Erbe rekom...

Page 116: ...tt desinfektionsmedel som r f renligt med reng ringsmedlet och som r l mpligt f r immersion Reng rings och desinfektionsmedel m ste vara l mpliga f r medicinprodukter av plast eller metall och ha ett...

Page 117: ...rodukten med filtrerad tryckluft 5 9 Maskinell reng ring och desinfektion Reng rings och desinfektionsapparatens effektivitet m ste i princip testas noggrant till exempel att CE m rkningen motsva rar...

Page 118: ...2 Kontrollera produktens mekaniska funktionsduglighet G r det att trycka p alla knappar Anv nd inte produkten om den r skadad 5 11 F rpacka 1 F rpacka produkten i en eng ngssteriliseringsf rpackning e...

Page 119: ...l 135 C Steriliseringsapparat enligt g llande nationella normer och f reskrifter till exempel DIN EN 13060 eller DIN EN 285 Steriliseringsprocess godk nd enligt DIN EN ISO 17665 F lj sterilisatortillv...

Page 120: ...120 OBJ_DOKU 182711 004 fm Seite 120 Dienstag 22 November 2016 3 25 15...

Page 121: ...FI OHJEITA K YTT VARTEN Hermotesti elektrodikahva 20190 148 OBJ_DOKU 182711 004 fm Seite 121 Dienstag 22 November 2016 3 25 15...

Page 122: ...4 Tuotteen liitt minen 125 5 Puhdistus desinfektio sterilointi 126 5 1 Turvaohjeita 126 5 2 Puhdistuksen rajoitukset 126 5 3 Suositeltu varustus aine 126 5 4 Suositellut toimenpiteet 127 5 5 Purkamin...

Page 123: ...tt vain yhdess hopeoitu jen palloelektrodien kanssa tuotenro 20191 079 tai 21191 052 Hermotesti elektrodikahva on ainoastaan tarkoitettu liitett v k si seuraaviin Erbe laitteisiin ICC 350 ZMK yhdess k...

Page 124: ...ista tarkastamista ennen jokaista k ytt Ohje Mik li kaapelin johdin on murtunut tai virrankulku kaape lissa on muulla tavoin katkennut saattaa paluuvirran yhteydess tai aktiivisessa virtapiiriss synty...

Page 125: ...yt sit 4 3 Tuotteen asentaminen Kiinnit hopeoitu palloelektrodi tuotenro 20191 079 tai 21191 052 hermotesti elektrodikahvaan tuotenro 20190 148 4 4 Tuotteen liitt minen Liit nt ICC 350 ZMK laitteesee...

Page 126: ...eriloinnissa ei saa ylitt 138 C n l mp tilaa l k yt kuumailmasterilointia 5 2 Puhdistuksen rajoitukset Toistuvalla uudelleenk ytt n valmistelulla on vaikutuksia t h n tuotteeseen Noudata tuotteen tark...

Page 127: ...valmistelumenetelm Muut poikkeavat mutta vastaavat menetelm t ovat mahdollisia mik li niit ei ole erityisesti kielletty K ytetyn menetelm n sopi vuuden varmistaminen asianmukaisilla toimenpiteill esi...

Page 128: ...ltava 5 5 11 Puhdistus ja desinfiointiaineen valmistajan ohjeita on ehdotto masti noudatettava 1 Valmista puhdistushaude ja erillinen desinfiointihaude val mistajan antamien ohjeiden mukaan 2 Upota tu...

Page 129: ...lle ja aseta tuote varovasti asian mukaiseen huuhtelukoriin Varmista ettei tuote kosketa muita instrumentteja instrumenttiosia 2 K ynnist tarkistettu ohjelma jolla on seuraavat ominaisuu det L mp desi...

Page 130: ...yrysterilointia seuraavassa kuvatulla menetelm ll Muiden sterilointimenetelmien k yt st Erbe Elektromedizin ei ota mink nlaista vastuuta H yrysterilointi Fraktioitu tyhji menetelm ja riitt v tuotteen...

Page 131: ...PL WSKAZ WKA U YTKOWA Uchwyt elektrody neurotest 20190 148 OBJ_DOKU 182711 004 fm Seite 131 Dienstag 22 November 2016 3 25 15...

Page 132: ...ktu 135 4 4 Przy czanie produktu 135 5 Czyszczenie dezynfekcja sterylizacja 136 5 1 Wskaz wki dotycz ce bezpiecze stwa 136 5 2 Ograniczenia powt rnego wykorzystania 136 5 3 Zalecane wyposa enie rodki...

Page 133: ...ktrycznie pobudliwe takich jak np nerw twarzowy Uchwyt elektrody neurotest powinien by stosowany wy cznie w po czeniu z elektrodami ze srebrn kul nr art 20191 079 lub 21191 052 Uchwyt elektrody Nerven...

Page 134: ...skontrolowa jego uszkodzenia mechaniczne Erbe Elektromedizin zaleca przeprowadzenie kontroli przej cia elektrycznego przed ka dym zastosowaniem niniejszego pro duktu Wskaz wka W przypadku p kni cia y...

Page 135: ...t 4 2 Sprawdzenie produktu Przed ka dym u yciem nale y sprawdzi izolacj kabli pod cze niowych funkcje aktywacji produktu oraz skontrolowa jego uszkodzenia mechaniczne Nie u ywa produktu w przypadku st...

Page 136: ...kt w Maksymalne ci nienie powietrza podczas suszenia 2 bar Maksymalne ci nienie wody podczas p ukania 2 bar Podczas maszynowego czyszczenia dezynfekcji nie przekracza temperatury 95 C Nale y konieczni...

Page 137: ...ie maszynowe w dezynfektorze G 7836 CD Miele Cie KG G tersloh Niemcy rodkiem do czyszczenia neodisher mediclean forte Dr Weigelt GmbH Co KG Hamburg Niemcy zgodnie z zalecanym progra mem mycia 5 4 Zale...

Page 138: ...rzanie Nale y stosowa rodek dezynfekcyjny kompatybilny ze rod kiem czyszcz cym kt ry nadaje si do przygotowania k pieli przez zanurzanie rodek czyszcz cy i dezynfekcyjny musi by przeznaczony do wyrob...

Page 139: ...ro dukt nie powinien przy tym dotyka innych cz ci w k pieli Nale y przestrzega zalecanego czasu dzia ania 7 Przep uka powierzchnie steryln wod dejonizowan min 1 minuta 8 Produkt wysuszy przefiltrowany...

Page 140: ...Kontrola 1 Nale y sprawdzi produkt pod k tem widocznych uszkodze i zu ycia Uszkodzenia w obr bie produktu np p kni cia szorst ko powierzchni ubytki materia u Uszkodzenia izolacji produktu i lub przewo...

Page 141: ...alno ci ze strony firmy Erbe Elektromedizin Sterylizacja parowa Metoda pr ni frakcjonowanej z wystarczaj cym suszeniem produktu Czas ekspozycji 3 do 18 minut w temperaturze 132 do 135 C Sterylizator z...

Page 142: ...142 OBJ_DOKU 182711 004 fm Seite 142 Dienstag 22 November 2016 3 25 15...

Page 143: ...CS POKYNY K POU IT Dr k elektrody Nervtest 20190 148 OBJ_DOKU 182711 004 fm Seite 143 Dienstag 22 November 2016 3 25 15...

Page 144: ...ku 147 4 4 P ipojen v robku 147 5 i t n dezinfekce sterilizace 148 5 1 Bezpe nostn pokyny 148 5 2 Omezen platn pro i t n 148 5 3 Doporu en vybaven prost edky 148 5 4 Doporu en postupy 149 5 5 Demont 1...

Page 145: ...s facialis Dr k elektrody Nervtest se sm pou vat pouze v kombinaci se st brnou kulovou elektrodou v r 20191 079 nebo 21191 052 Dr k elektrody Nervtest je ur en k p ipojen v lu n k n sledu j c m p stro...

Page 146: ...zkou ku elektrick pr chodnosti Upozorn n Pokud je kabelov vodi zlomen nebo je elektrick pr chodnost kabelu jinak p eru en m e doj t ve zp tn m vo di i nebo v aktivn m proudov m okruhu k vytvo en elekt...

Page 147: ...ek nepou vejte 4 3 Mont v robku Zasu te st brnou kulovou elektrodu v r 20191 079 nebo 21191 052 do dr ku elektrody Nervtest v r 20190 148 4 4 P ipojen v robku P ipojen k ICC 350 ZMK Dr k zasu t do zd...

Page 148: ...pro i t n ast op tn p prava m vliv na v robek Dodr ujte bezpe nostn pokyny ke kontrole v robku P i zjevn m po kozen nebo naru en funk nosti se v robek ji nesm pou vat 5 3 Doporu en vybaven prost edky...

Page 149: ...ji t n spolehlivosti skute n pou it ch postup dodr en m p slu n ch opat en nap validace rutinn kontrola kontrola kompatibility materi l 5 5 Demont 1 Vyjm te VF elektrodu z rukojeti 5 6 Pot ebn pomocn...

Page 150: ...z um l hmoty a kovu kter maj hod notu pH mezi 5 5 a 11 Bezpodm ne n dodr ujte daje uveden v robcem istic ch a dezinfek n ch prost edk 1 Podle daj v robce p ipravte ist c l ze a zvl t dezin fek n l ze...

Page 151: ...te daje uveden v robcem istic ch a dezinfek n ch prost edk 1 V robek opatrn vlo te s voln svinut m kabelem do vhod n ho om vac ho ko e Dbejte na to aby se v robek nedot kal dn ch dal ch p stroj nebo j...

Page 152: ...ru 5 12 Sterilizace Sterilizujte pouze ist a dezinfikovan v robky Firma Erbe Elektromedizin doporu uje parn sterilizaci n e po psanou metodou Pou it jin metody sterilizace je mimo odpo v dnost firmy E...

Page 153: ...HU ALKALMAZ SI TMUTAT Nervtest elektr dafog 20190 148 OBJ_DOKU 182711 004 fm Seite 153 Dienstag 22 November 2016 3 25 15...

Page 154: ...4 A term k csatlakoztat sa 157 5 Tiszt t s fert tlen t s steriliz ci 158 5 1 Biztons gi utas t sok 158 5 2 Az jrafelhaszn l s korl toz sa 158 5 3 Aj nlott felszerel s eszk z k 159 5 4 Aj nlott elj r s...

Page 155: ...aliz l s ra val A Nervtest elektr dafog csak az ez st g mbelektr d kkal cikk sz m 20191 079 vagy 21191 052 kombin lva haszn lhat A Nervtest elektr dafog kiz r lag az al bbi Erbe k sz l kekhez t rt n c...

Page 156: ...s hogy a term knek van e mechanikai s r l se Az Erbe Elektromedizin GmbH e term k minden egyes haszn la ta el tt javasolja az elektromos raml s vizsg lat t Megjegyz s Amennyiben a k belvezet k meg va...

Page 157: ...steriliz lni kell 4 2 A term k ellen rz se Minden egyes haszn lat el tt ellen rizze le a csatlakoz k belek szigetel s t a term k m k d si funkci it s hogy a term knek van e mechanikai s r l se Amennyi...

Page 158: ...t n j l le kell mosni Maxim lis leveg nyom s sz r t skor 2 bar Maxim lis v znyom s bl t skor 2 bar G pi tiszt t s fert tlen t s sor n tilos a 95 C os h m rs kletet t ll pni A tiszt t s fert tlen t esz...

Page 159: ...tlen t k sz l kben neodisher mediclean forte Dr Weigert GmbH Co KG Hamburg N metorsz g tiszt t szerrel az aj nlott mos program szerint 5 4 Aj nlott elj r sok Az Erbe az al bbi el k sz t si elj r st ja...

Page 160: ...d k sz t s hez megfelel a tiszt t szerrel kompatibilis fert tlen t szert A tiszt t s a fert tlen t szer alkalmas kell legyen m anyagb l s f mb l k sz lt orvosi eszk z k n val haszn latra s 5 5 s 11 k...

Page 161: ...t id t 7 bl tse le a fel leteket steril ionmentes v zzel legal bb 1 percig 8 Sz r tsa meg a term ket sz rt s r tett leveg vel 5 9 G pi tiszt t s s fert tlen t s Csak olyan tiszt t s fert tlen t k sz l...

Page 162: ...1 Ellen rizze a term ket a szemmel szrevehet s r l sek s kop sok szempontj b l A term ken tal lhat s r l sek pl reped sek rdes fe l let lepattogz d sok A term k s vagy a k bel dug s csatlakoz szigete...

Page 163: ...get G zsteriliz ci Frakcion lt v kuumos elj r s a term k megfelel sz r t s val Kezel si id 3 18 perc 132 s 135 C k z tti h m rs k leten Az rv nyes orsz gos norm knak s el r soknak megfelel steriliz to...

Page 164: ...164 OBJ_DOKU 182711 004 fm Seite 164 Dienstag 22 November 2016 3 25 15...

Page 165: ...RU 20190 148 OBJ_DOKU 182711 004 fm Seite 165 Dienstag 22 November 2016 3 25 15...

Page 166: ...7 3 168 4 169 4 1 169 4 2 169 4 3 169 4 4 170 5 170 5 1 170 5 2 171 5 3 171 5 4 171 5 5 172 5 6 172 5 7 172 5 8 173 5 9 174 5 10 174 5 11 175 5 12 175 OBJ_DOKU 182711 004 fm Seite 166 Dienstag 22 Nove...

Page 167: ...167 Erbe 1 20191 079 21191 052 Erbe ICC 350 ZMK FB2 20121 007 NT 2 NT2 20142 200 NT2 20142 100 20142 101 2 Vp OBJ_DOKU 182711 004 fm Seite 167 Dienstag 22 November 2016 3 25 15...

Page 168: ...168 3 Erbe Elektromedizin OBJ_DOKU 182711 004 fm Seite 168 Dienstag 22 November 2016 3 25 15...

Page 169: ...169 Erbe Elektromedizin Erbe Elektromedizin 4 4 1 4 2 4 3 20191 079 21191 052 20190 148 OBJ_DOKU 182711 004 fm Seite 169 Dienstag 22 November 2016 3 25 15...

Page 170: ...170 4 4 ICC 350 ZMK Erbe ICC 350 ZMK Erbe ICC 350 ZMK NT 2 NT 2 5 5 1 2 2 95 C OBJ_DOKU 182711 004 fm Seite 170 Dienstag 22 November 2016 3 25 15...

Page 171: ...3 Cidezyme Cidex OPA Johnson Johnson Medical Limited Gargrave Skipton G 7836 CD Miele Cie KG neodisher mediclean forte Dr Weigert GmbH Co KG 5 4 Erbe OBJ_DOKU 182711 004 fm Seite 171 Dienstag 22 Novem...

Page 172: ...172 5 5 1 5 6 5 7 1 2 OBJ_DOKU 182711 004 fm Seite 172 Dienstag 22 November 2016 3 25 15...

Page 173: ...173 5 8 pH 5 5 11 1 2 3 4 1 5 6 7 1 8 OBJ_DOKU 182711 004 fm Seite 173 Dienstag 22 November 2016 3 25 15...

Page 174: ...174 5 9 CE DIN EN ISO 15883 pH 5 5 11 1 2 5 10 90 93 C A0 3000 3 5 10 1 OBJ_DOKU 182711 004 fm Seite 174 Dienstag 22 November 2016 3 25 15...

Page 175: ...175 2 5 11 1 5 12 Erbe Elektromedizin Erbe Elektromedizin 3 18 132 135 C DIN EN 13060 DIN EN 285 DIN EN ISO 17665 OBJ_DOKU 182711 004 fm Seite 175 Dienstag 22 November 2016 3 25 15...

Page 176: ...176 OBJ_DOKU 182711 004 fm Seite 176 Dienstag 22 November 2016 3 25 15...

Page 177: ...TR KULLANIM TALIMATI Sinir Testi i in Elektrot Tutaca 20190 148 OBJ_DOKU 182711 004 fm Seite 177 Dienstag 22 November 2016 3 25 15...

Page 178: ...181 5 Temizleme dezenfeksiyon sterilizasyon 182 5 1 G venlik notlar 182 5 2 Yeniden kullan ma haz rlama s n rl l 182 5 3 Tavsiye edilen aksesuar ajanlar 182 5 4 Tavsiye edilen y ntemler 183 5 5 Par al...

Page 179: ...syonu i in tasarlanm t r Sinir testi elektrot tutaca yaln zca g m top elektrotlarla r n no 20191 079 veya 21191 052 birlikte kullan labilir Sinir testi elektrot tutaca yaln zca u Erbe cihazlar na ba l...

Page 180: ...sinde kontrol edilmesini nermektedir Not E er kablodaki iletken k r l r ya da kablonun elektriksel ge irgenli i herhangi bir ekilde kesintiye u rarsa hastadan d nen kabloda veya aktif ak m devresinde...

Page 181: ...aj G m top elektrodu r n no 20191 079 veya 21191 052 si nir testi elektrot tutaca n n r n no 20190 148 i ine sokun 4 4 r n g ebekesine ba lay n ICC 350 ZMK ye ba lama Tutaca Erbe ICC 350 ZMK nin Neuro...

Page 182: ...ilizasyon i lemlerinde 138 C yi a may n S cak hava ile sterilize etmeyin 5 2 Yeniden kullan ma haz rlama s n rl l S k olarak yeniden kullan ma haz rlanmas bu r n zerinde etkili olur r n n kontrol yle...

Page 183: ...tavsiye ediyoruz A k a ekarte edilmi olmamas art yla e de er farkl y ntemler de uygulanabilir Ger ekten uygulanan y ntemin el veri li olup olmad n n uygun nlemlerle rn validasyon rutin denetimler mal...

Page 184: ...enfeksiyon maddesi reticilerinin talimatlar na mutlaka uyulmal d r 1 retici talimatlar na g re bir temizleme banyosu ve ayr bir dezenfeksiyon banyosu haz rlay n 2 r n banyodaki di er par alara temas e...

Page 185: ...emas etmemesine dikkat edin 2 A a daki zelliklere sahip denetlenmi bir program ba la t n Termal dezenfeksiyon 90 ila 93 C de 5 ila 10 dakika A0 3000 Son durulama distile veya demineralize su ile yap l...

Page 186: ...asyon y ntemlerinin uygulanmas Erbe Elektromedizin firmas n n sorumlulu u d nda kal r Buharl sterilizasyon Yeterli kurutulmu r nle b l nm vakum y ntemi 132 ila 135 C de tutma s resi 3 ila 18 dakikad r...

Page 187: ...ZH 20190 148 OBJ_DOKU 182711 004 fm Seite 187 Dienstag 22 November 2016 3 25 15...

Page 188: ...9 3 189 4 190 4 1 190 4 2 190 4 3 191 4 4 191 5 191 5 1 191 5 2 192 5 3 192 5 4 192 5 5 192 5 6 193 5 7 193 5 8 193 5 9 194 5 10 194 5 11 195 5 12 195 OBJ_DOKU 182711 004 fm Seite 188 Dienstag 22 Nove...

Page 189: ...189 Erbe 1 20191 079 21191 052 Erbe ICC 350 ZMK FB2 20121 007 NT 2 NT2 20142 200 NT2 20142 100 20142 101 2 Vp 3 OBJ_DOKU 182711 004 fm Seite 189 Dienstag 22 November 2016 3 25 15...

Page 190: ...190 Erbe Elektromedizin Erbe Erbe 4 4 1 4 2 OBJ_DOKU 182711 004 fm Seite 190 Dienstag 22 November 2016 3 25 15...

Page 191: ...191 4 3 20191 079 21191 052 20190 148 4 4 ICC 350 ZMK Erbe ICC 350 ZMK Erbe ICC 350 ZMK 1 5 NT 2 NT 2 5 5 1 2 bar 2 bar 95 138 OBJ_DOKU 182711 004 fm Seite 191 Dienstag 22 November 2016 3 25 15...

Page 192: ...A Johnson Johnson Medical Limited Gargrave Skipton G 7836 CD Miele Cie KG Guetersloh Germany neodisher forte Dr Weigert GmbH Co KG Hamburg Germany 5 4 Erbe 5 5 1 OBJ_DOKU 182711 004 fm Seite 192 Diens...

Page 193: ...193 5 6 5 7 1 5 8 pH 5 5 11 1 2 3 4 1 5 6 2 bar OBJ_DOKU 182711 004 fm Seite 193 Dienstag 22 November 2016 3 25 15...

Page 194: ...194 7 1 8 5 9 DIN EN ISO 15883 CE pH 5 5 11 1 2 90 93 5 10 A0 3000 3 5 10 1 2 OBJ_DOKU 182711 004 fm Seite 194 Dienstag 22 November 2016 3 25 15...

Page 195: ...195 5 11 1 5 12 Erbe Erbe 132 135 3 18 DIN EN 13060 DIN EN 285 DIN EN ISO 17665 OBJ_DOKU 182711 004 fm Seite 195 Dienstag 22 November 2016 3 25 15...

Page 196: ...196 OBJ_DOKU 182711 004 fm Seite 196 Dienstag 22 November 2016 3 25 15...

Page 197: ...OBJ_DOKU 182711 004 fm Seite 197 Dienstag 22 November 2016 3 25 15...

Page 198: ...Belgi Belgique Belgien Zaventem Tel 32 2 254 88 60 info erbe belgium com 86 21 62758 440 info erbechina com France Limonest T l 33 4 78 64 92 55 erbe erbe france com India Chennai Tel 91 44 7155 5100...

Reviews: