WARNING!
3
Safety instructions
This product may be used only by trained medical staff, who
have been shown how to use it according to the instructions for
use.
Not intended for direct contact with the heart or the central cir-
culatory system!
In the case of patients wearing cardiac pacemakers or other ac-
tive implants there is a potential risk due to the fact that there
may be interference with pacemaker operation or the pacemaker
may be damaged. In case of doubt obtain expert advice.
Before every use check the insulation of the connecting cables,
the activation functions of the product and whether the product
has any physical damage.
Erbe Elektromedizin recommends performing an electrical con-
tinuity test before every use of this product.
Note: If the cable conductor is broken or the electrical continuity
of the cable is interrupted in any other manner it can cause arc-
ing in the patient return system or in the active circuit; this can
lead to the patient suffering burns or the outbreak of a fire.
If damaged, do not use this product!
Do not use in the presence of combustible or explosive materi-
als!
Never lay this product on the patient or in his/her direct vicinity.
Protect this product from any form of mechanical damage! Do
not throw! Do not use force! If there is a cable, and unless stated
otherwise, do not kink it or wrap it around the accompanying
product!
Erbe Elektromedizin expressly warns against modifying the
product. Any modification exempts Erbe Elektromedizin from
any and all liability.
OBJ_DOKU-182711-004.fm Seite 18 Dienstag, 22. November 2016 3:25 15
Summary of Contents for Nervtest 20190-148
Page 2: ...OBJ_DOKU 182711 004 fm Seite 2 Dienstag 22 November 2016 3 25 15...
Page 14: ...14 OBJ_DOKU 182711 004 fm Seite 14 Dienstag 22 November 2016 3 25 15...
Page 36: ...36 OBJ_DOKU 182711 004 fm Seite 36 Dienstag 22 November 2016 3 25 15...
Page 48: ...48 OBJ_DOKU 182711 004 fm Seite 48 Dienstag 22 November 2016 3 25 15...
Page 60: ...60 OBJ_DOKU 182711 004 fm Seite 60 Dienstag 22 November 2016 3 25 15...
Page 72: ...72 OBJ_DOKU 182711 004 fm Seite 72 Dienstag 22 November 2016 3 25 15...
Page 73: ...EL 20190 148 OBJ_DOKU 182711 004 fm Seite 73 Dienstag 22 November 2016 3 25 15...
Page 76: ...76 3 Erbe Elektromedizin OBJ_DOKU 182711 004 fm Seite 76 Dienstag 22 November 2016 3 25 15...
Page 80: ...80 5 5 1 5 6 5 7 1 5 8 2 bar OBJ_DOKU 182711 004 fm Seite 80 Dienstag 22 November 2016 3 25 15...
Page 84: ...84 OBJ_DOKU 182711 004 fm Seite 84 Dienstag 22 November 2016 3 25 15...
Page 96: ...96 OBJ_DOKU 182711 004 fm Seite 96 Dienstag 22 November 2016 3 25 15...
Page 108: ...108 OBJ_DOKU 182711 004 fm Seite 108 Dienstag 22 November 2016 3 25 15...
Page 120: ...120 OBJ_DOKU 182711 004 fm Seite 120 Dienstag 22 November 2016 3 25 15...
Page 142: ...142 OBJ_DOKU 182711 004 fm Seite 142 Dienstag 22 November 2016 3 25 15...
Page 164: ...164 OBJ_DOKU 182711 004 fm Seite 164 Dienstag 22 November 2016 3 25 15...
Page 165: ...RU 20190 148 OBJ_DOKU 182711 004 fm Seite 165 Dienstag 22 November 2016 3 25 15...
Page 168: ...168 3 Erbe Elektromedizin OBJ_DOKU 182711 004 fm Seite 168 Dienstag 22 November 2016 3 25 15...
Page 172: ...172 5 5 1 5 6 5 7 1 2 OBJ_DOKU 182711 004 fm Seite 172 Dienstag 22 November 2016 3 25 15...
Page 176: ...176 OBJ_DOKU 182711 004 fm Seite 176 Dienstag 22 November 2016 3 25 15...
Page 187: ...ZH 20190 148 OBJ_DOKU 182711 004 fm Seite 187 Dienstag 22 November 2016 3 25 15...
Page 196: ...196 OBJ_DOKU 182711 004 fm Seite 196 Dienstag 22 November 2016 3 25 15...
Page 197: ...OBJ_DOKU 182711 004 fm Seite 197 Dienstag 22 November 2016 3 25 15...