Cleaning, disinfection, sterilization
5.1 Safety
instructions
May be cleaned in an ultrasonic bath.
Never use sharp objects for cleaning.
Disinfectants must be rinsed off thoroughly after use.
Maximum air pressure during drying: 2 bar.
Maximum water pressure during rinsing: 2 bar.
During machine cleaning and disinfection, do not exceed 95 °C.
It is essential to follow the cleaning agent and disinfectant man-
ufacturers’ instructions.
During sterilization, do not exceed 138 °C.
Do not sterilize in hot air.
5.2 Reconditioning
limitation
Frequent reconditioning can damage this product. Observe the
safety instructions for checking the product. With visible damage
or functional impairment, the product may no longer be used.
5.3 Recommended equipment / substances
The service life of the product, cleaning action and sterility are
validated for the following cleaning and preparation procedures
if the recommended or equivalent substances are used in accor-
dance with the manufacturer's instructions.
•
Manual cleaning/disinfection with the detergent
Cidezyme ® and disinfectant Cidex ® OPA (Johnson & John-
son Medical Limited, Gargrave, Skipton).
•
Machine cleaning and disinfection in a G 7836 CD disinfector
(Miele & Cie. KG, Gütersloh, Germany) with the cleaning
OBJ_DOKU-182711-004.fm Seite 20 Dienstag, 22. November 2016 3:25 15
Summary of Contents for Nervtest 20190-148
Page 2: ...OBJ_DOKU 182711 004 fm Seite 2 Dienstag 22 November 2016 3 25 15...
Page 14: ...14 OBJ_DOKU 182711 004 fm Seite 14 Dienstag 22 November 2016 3 25 15...
Page 36: ...36 OBJ_DOKU 182711 004 fm Seite 36 Dienstag 22 November 2016 3 25 15...
Page 48: ...48 OBJ_DOKU 182711 004 fm Seite 48 Dienstag 22 November 2016 3 25 15...
Page 60: ...60 OBJ_DOKU 182711 004 fm Seite 60 Dienstag 22 November 2016 3 25 15...
Page 72: ...72 OBJ_DOKU 182711 004 fm Seite 72 Dienstag 22 November 2016 3 25 15...
Page 73: ...EL 20190 148 OBJ_DOKU 182711 004 fm Seite 73 Dienstag 22 November 2016 3 25 15...
Page 76: ...76 3 Erbe Elektromedizin OBJ_DOKU 182711 004 fm Seite 76 Dienstag 22 November 2016 3 25 15...
Page 80: ...80 5 5 1 5 6 5 7 1 5 8 2 bar OBJ_DOKU 182711 004 fm Seite 80 Dienstag 22 November 2016 3 25 15...
Page 84: ...84 OBJ_DOKU 182711 004 fm Seite 84 Dienstag 22 November 2016 3 25 15...
Page 96: ...96 OBJ_DOKU 182711 004 fm Seite 96 Dienstag 22 November 2016 3 25 15...
Page 108: ...108 OBJ_DOKU 182711 004 fm Seite 108 Dienstag 22 November 2016 3 25 15...
Page 120: ...120 OBJ_DOKU 182711 004 fm Seite 120 Dienstag 22 November 2016 3 25 15...
Page 142: ...142 OBJ_DOKU 182711 004 fm Seite 142 Dienstag 22 November 2016 3 25 15...
Page 164: ...164 OBJ_DOKU 182711 004 fm Seite 164 Dienstag 22 November 2016 3 25 15...
Page 165: ...RU 20190 148 OBJ_DOKU 182711 004 fm Seite 165 Dienstag 22 November 2016 3 25 15...
Page 168: ...168 3 Erbe Elektromedizin OBJ_DOKU 182711 004 fm Seite 168 Dienstag 22 November 2016 3 25 15...
Page 172: ...172 5 5 1 5 6 5 7 1 2 OBJ_DOKU 182711 004 fm Seite 172 Dienstag 22 November 2016 3 25 15...
Page 176: ...176 OBJ_DOKU 182711 004 fm Seite 176 Dienstag 22 November 2016 3 25 15...
Page 187: ...ZH 20190 148 OBJ_DOKU 182711 004 fm Seite 187 Dienstag 22 November 2016 3 25 15...
Page 196: ...196 OBJ_DOKU 182711 004 fm Seite 196 Dienstag 22 November 2016 3 25 15...
Page 197: ...OBJ_DOKU 182711 004 fm Seite 197 Dienstag 22 November 2016 3 25 15...