5.4 Recommended
methods
Erbe recommends the preparation procedures described below.
Equivalent different procedures are possible if not explicitly ex-
cluded. It is incumbent on the user to ensure the suitability of the
actual procedures used by means of suitable measures (e.g. val-
idation, routine monitoring, check of material compatibility).
5.5 Dismantling
1. Remove the HF electrode from the handle.
2. For conization electrode Art. No. 20191-075 only: Remove
the loop out of the conization electrode.
5.6 Required
Aids
5.7 Precleaning
Use water for precleaning, a non-fixing disinfectant if necessary.
1. Remove surface contaminants with a soft brush/cloth while
holding the product in a water bath and/or rinsing the prod-
uct under running water.
Reprocessing step Items Needed
Precleaning
Soft plastic brush/soft cloth
OBJ_DOKU-153232-006.fm Seite 19 Dienstag, 22. November 2016 11:16 11
Summary of Contents for 20191-075
Page 2: ...OBJ_DOKU 153232 006 fm Seite 2 Dienstag 22 November 2016 11 16 11...
Page 22: ...22 OBJ_DOKU 153232 006 fm Seite 22 Dienstag 22 November 2016 11 16 11...
Page 32: ...32 OBJ_DOKU 153232 006 fm Seite 32 Dienstag 22 November 2016 11 16 11...
Page 52: ...52 OBJ_DOKU 153232 006 fm Seite 52 Dienstag 22 November 2016 11 16 11...
Page 92: ...92 OBJ_DOKU 153232 006 fm Seite 92 Dienstag 22 November 2016 11 16 11...
Page 102: ...102 OBJ_DOKU 153232 006 fm Seite 102 Dienstag 22 November 2016 11 16 11...
Page 112: ...112 OBJ_DOKU 153232 006 fm Seite 112 Dienstag 22 November 2016 11 16 11...
Page 132: ...132 OBJ_DOKU 153232 006 fm Seite 132 Dienstag 22 November 2016 11 16 11...
Page 162: ...162 OBJ_DOKU 153232 006 fm Seite 162 Dienstag 22 November 2016 11 16 11...
Page 171: ...OBJ_DOKU 153232 006 fm Seite 171 Dienstag 22 November 2016 11 16 11...