6)
Enter Menu 23, select transducer type
“5. Plug-in B45”
;
7)
Enter Menu 24, select transducer-mounting method
“1. Z-method”
;
8)
Enter Menu 01 to display measurement result. At this moment, the S and Q value should
be 0 in Menu 01.
3.
Transducers Operation
Put a little water or couplant on measuring plane of transducers, put the measuring parts of
two transducers face to face closely, shown as below.
4.
How to judge the flow meter system
Keep the transducers status shown as above, and then view the S and Q value in Menu 01.
1)
If the two S values are larger then 60, and the difference of the two values is smaller than
10, it means the system is working well.
2)
If the two S values have a great difference which is larger than 10, or there is one S value
is 0, it means the wirings or transducers have problem.
Check the wirings. If the wirings are OK, customers need replace the transducers or send
them back for repair.
3)
If the two S values are both 0, it means the transmitter or transducers have problem.
Check the wirings, if the wirings are OK, customers need replace the meter or send it back
for repair.
How to Judge the Liquid Flowing Direction
Make sure that the instrument works properly;
Check the flow rate for the indication. If the displayed value is positive, the direction of the
flow will be from the UP transducer to the Down transducer; if the displayed value is negative,
the direction will be from the Down transducer to the UP transducers;
How to Reset the Default Setups
Use M37, it has another function to recover the default setups. When the ‘selection’ message is
displayed, press the dot key
.
first, then press key
◀
, meter will erase all the parameters
entered by the user and setup the meter with default values.
35
ULTRA-E-FB&C
Operation of Insertion Transducers
Operation of Clamp-on Transducers