Performing a Blood Glucose Test with EasyMax 15 Test Strips

Please adhere to your institution’s medical device handling and site preparation procedures.

1.  Wash your patient’s hands 

  with soap and warm water. 

  Rinse and dry thoroughly.

  Your institution may require 

  that you use an alcohol 

  prep pad. Please ensure

  that the test site is 

  completely dry before 

  proceeding. 

2.  Put a test strip into the meter in 

  the direction of the arrow. The 

  meter turns on and the icon 

  shows automatically.

3.  Press    (Left Button) or

  (Right Button) to set a 

  Patient Number and press  

  to confirm the setting (only if   

        function if on).

4.  Please wait at least 5 

  seconds until the intended 

  area is dry and clean before 

  using the safety lancet.

5.  Remove the protective lancet 

  cap.

6.  Position the lancet and press 

  it firmly against the puncture 

  site.

7. Gently squeeze and/or 

  massage the fingertip until 

  a round drop of blood forms

  on the fingertip.

 

Page 9

Note:  If the function of      is on, and you do not choose a medical record number before applying 

    blood, the test result will be recorded under 

        .  .

9.  The meter starts to count 

  down from 5 seconds and 

  then displays the test result.

 

8. Touch the blood drop at the tip of

  the transparent window of the test

  strip. 

Do not put blood on top of 

  the strip.

 Be sure to get enough

  blood on the strip’s reaction zone. 

  Otherwise, an error code will 

  appear.

 

10.  Push the Strip Ejector to eject the test strip. Properly dispose of both the 

 

strip and the lancet.

11.  After discarding, wash hands thoroughly with soap and water. Rinse and dry

 

thoroughly.

12.  Clean and disinfect the meter following the instructions on page 4.

13.  Change gloves between patients.

In a Long Term Care facility setting, we recommend that a safety lancet be used for each individual glucose 

test.

Summary of Contents for EasyMax LTC

Page 1: ...ng Blood Glucose System This Manual is intended for facilities where point of care testing is conducted by a healthcare professional Please thoroughly read the EasyMax LTC Owner s Guide and all produc...

Page 2: ...tured by EPS Bio Technology Corp No 8 R D Road III Hsinchu Science Park Hsinchu Taiwan 30077 Customer Service 1 866 994 3345 Toll Free Mon Fri 8 00 AM 6 00 PM EasyMax LTC Self Monitoring Blood Glucose...

Page 3: ...Meter is designed and approved for testing fresh capillary whole blood samples from your fingertip palm and forearm The meter is for outside the body in vitro use It should not be used to diagnose di...

Page 4: ...ic Acid 0 8 1 2 4 4 Dopamine NA 0 09 2 6 Gentisic Acid 1 8 1 8 6 Glipizide 0 2 0 2 8 Ibuprofen 0 5 4 2 50 50 L Dopa 0 5 8 0 2 Methyldopa 0 1 0 5 0 5 2 Sodium Salicylate 15 30 50 Tetracycline 0 4 1 5 1...

Page 5: ...ts If the results are inconsistent with your patient s symptoms please follow these steps 1 Run a control solution test page 7 2 Repeat a blood glucose test page 9 3 After running the quality control...

Page 6: ...rmicidal wipe Open unfold and use the first germicidal wipe to remove heavy soil Step 2 Disinfection Instruction After cleaning and before using the meter again unfold a new wipe and thoroughly wet al...

Page 7: ...st result or error message E Insert a Test Strip Insert a test strip to start testing blood glucose F Medical Record No Shows the patient number G Apply Control Solution Control solution test marker H...

Page 8: ...e Use Setting the current time and date on your meter is important if you use the meter memory 4 Repeat step 3 to set the date and time The flashing field is the one you are currently setting 5 Press...

Page 9: ...mber you select if you flag the result as a control solution test The result will be displayed in the meter s memory as a control solution test Page 7 2 Press Left Button to select the mode of Control...

Page 10: ...ol test with properly stored strips Was the test strip bottle capped tightly If it wasn t the strips could be humidified Replace the test strips Is the control solution expired or contaminated Replace...

Page 11: ...firmly against the puncture site 7 Gently squeeze and or massage the fingertip until a round drop of blood forms on the fingertip Page 9 Note If the function of is on and you do not choose a medical...

Page 12: ...f 500 F 1040 F 100 C 400 C This alerts users that an incorrect result may occur if the test continues Action Relocate the meter to a location with temperature between 500 F 1040 F 100 C 400 C The LF a...

Page 13: ...EasyMax LTC Self Monitoring Blood Glucose System Training Information for Healthcare Professionals Page 11...

Page 14: ...end of test strip Identifies the designated hand hold area 4 Demonstrates proper test strip use Only uses EasyMax LTC blood glucose meters with EasyMax 15 blood glucose test strips Runs a control solu...

Page 15: ...Quality Control Performs a control solution test and can explain why it should be done Understands the dates listed on the control solution and how they apply to opened vs unopened vials Recognizes wh...

Page 16: ...ill lower or higher than expected D All of the Above 4 When performing a blood glucose test where should you apply the blood A On top of the test strip B On the side of the test strip C The tip of the...

Page 17: ...type of battery does the EasyMax LTC require A 3 Volt Lithium B Alkaline 1 5 Volt AAA C Alkaline 1 5 Volt AAAA D None of the above 12 Describe the proper way to dispose of a safety lancet 13 Describe...

Page 18: ...Glucose Quality Control Results Log Facility Name Monitor Month Year Results Reviewed by Date Time Staff Performing Test Strip Lot Level 1 Control Level 2 Control Action Taken Lot Range Result Lot Ran...

Reviews: