5.  Demonstration of blood glucose testing.
 

 

 Demonstrates proper safety lancet technique when collecting blood sample

 

 

 Demonstrates appropriate disposal of safety lancets in accordance with institution’s policy

 

 

 Demonstrates the correct way to insert the test strip into meter

 

 

 Understands symbols that appear on the meter screen and what, if any, corresponding 

 

 

 action

 

 

 Demonstrates correct placement of blood on test strip

 

 

 Records results as specified per your institution

 

 

 Understands the protocol for inaccurate results (re-tests, control solution test, 

 

 

 troubleshoot)

 

 

 Knows the actions to be taken when results are abnormally high or low

6.  Quality Control

 

 

 Performs a control solution test and can explain why it should be done

 

 

 Understands the dates listed on the control solution and how they apply to opened vs 

 

 

 unopened vials

 

 

 Recognizes what it means when a control solution test result falls in the range listed on

 

 

 the strip vial, and when the test result falls out of the range on the vial

7.  Troubleshooting Guidelines
 

 

 Understands the steps to follow when troubleshooting

 

 

 Knows the warranty of meter

 

 

 Knows how to handle defective meter, strips, etc. (next steps, manufacturer customer 

 

 

 service phone number)

8.  Proper Care
 

 

 Understands how to maintain the meter

 

 

 Locates proper storage

 

 

 Demonstrates proper cleaning

 

 

 Demonstrates proper battery installation

9.  Documentation
 

 

 Understands what information must be documented by law and what is required by your

 

 

 institution.

Page 13

Summary of Contents for EasyMax LTC

Page 1: ...ng Blood Glucose System This Manual is intended for facilities where point of care testing is conducted by a healthcare professional Please thoroughly read the EasyMax LTC Owner s Guide and all produc...

Page 2: ...tured by EPS Bio Technology Corp No 8 R D Road III Hsinchu Science Park Hsinchu Taiwan 30077 Customer Service 1 866 994 3345 Toll Free Mon Fri 8 00 AM 6 00 PM EasyMax LTC Self Monitoring Blood Glucose...

Page 3: ...Meter is designed and approved for testing fresh capillary whole blood samples from your fingertip palm and forearm The meter is for outside the body in vitro use It should not be used to diagnose di...

Page 4: ...ic Acid 0 8 1 2 4 4 Dopamine NA 0 09 2 6 Gentisic Acid 1 8 1 8 6 Glipizide 0 2 0 2 8 Ibuprofen 0 5 4 2 50 50 L Dopa 0 5 8 0 2 Methyldopa 0 1 0 5 0 5 2 Sodium Salicylate 15 30 50 Tetracycline 0 4 1 5 1...

Page 5: ...ts If the results are inconsistent with your patient s symptoms please follow these steps 1 Run a control solution test page 7 2 Repeat a blood glucose test page 9 3 After running the quality control...

Page 6: ...rmicidal wipe Open unfold and use the first germicidal wipe to remove heavy soil Step 2 Disinfection Instruction After cleaning and before using the meter again unfold a new wipe and thoroughly wet al...

Page 7: ...st result or error message E Insert a Test Strip Insert a test strip to start testing blood glucose F Medical Record No Shows the patient number G Apply Control Solution Control solution test marker H...

Page 8: ...e Use Setting the current time and date on your meter is important if you use the meter memory 4 Repeat step 3 to set the date and time The flashing field is the one you are currently setting 5 Press...

Page 9: ...mber you select if you flag the result as a control solution test The result will be displayed in the meter s memory as a control solution test Page 7 2 Press Left Button to select the mode of Control...

Page 10: ...ol test with properly stored strips Was the test strip bottle capped tightly If it wasn t the strips could be humidified Replace the test strips Is the control solution expired or contaminated Replace...

Page 11: ...firmly against the puncture site 7 Gently squeeze and or massage the fingertip until a round drop of blood forms on the fingertip Page 9 Note If the function of is on and you do not choose a medical...

Page 12: ...f 500 F 1040 F 100 C 400 C This alerts users that an incorrect result may occur if the test continues Action Relocate the meter to a location with temperature between 500 F 1040 F 100 C 400 C The LF a...

Page 13: ...EasyMax LTC Self Monitoring Blood Glucose System Training Information for Healthcare Professionals Page 11...

Page 14: ...end of test strip Identifies the designated hand hold area 4 Demonstrates proper test strip use Only uses EasyMax LTC blood glucose meters with EasyMax 15 blood glucose test strips Runs a control solu...

Page 15: ...Quality Control Performs a control solution test and can explain why it should be done Understands the dates listed on the control solution and how they apply to opened vs unopened vials Recognizes wh...

Page 16: ...ill lower or higher than expected D All of the Above 4 When performing a blood glucose test where should you apply the blood A On top of the test strip B On the side of the test strip C The tip of the...

Page 17: ...type of battery does the EasyMax LTC require A 3 Volt Lithium B Alkaline 1 5 Volt AAA C Alkaline 1 5 Volt AAAA D None of the above 12 Describe the proper way to dispose of a safety lancet 13 Describe...

Page 18: ...Glucose Quality Control Results Log Facility Name Monitor Month Year Results Reviewed by Date Time Staff Performing Test Strip Lot Level 1 Control Level 2 Control Action Taken Lot Range Result Lot Ran...

Reviews: