Page 1: ...MU Self Monitoring Blood Glucose System User s Manual Please read this User s Manual thoroughly before using your blood glucose meter...

Page 2: ...n thank you for choosing the EASYMAX MU SMBG System Intended Use The EASYMAX MU Self Monitoring Blood Glucose Test System is intended for the quantitative measurement of glucose in fresh capillary who...

Page 3: ...essories Blood Glucose Meter Alkaline Battery 2 ct Lancets 10 pcs Lancing Device User s Manual Warranty Card Carrying Case AST Lancing Device Cap Blood Glucose Test Strips 25 pcs Level 1 Control Solut...

Page 4: ...for your comfort Do you need help If you have questions or need assistance please contact your healthcare professional or visit EASYMAX Website www easymaxdiabetescare com for diabetes management tool...

Page 5: ...isassemble the meter as this may cause damage to the components resulting in incorrect readings Disassembling the meter will also void the warranty Always keep the meter clean and store it in a safe p...

Page 6: ...iohazard Healthcare professionals using this system on multiple patients should be aware that all products or objects that come in contact with human blood even after cleaning should be handled as if...

Page 7: ...t away If you have followed the steps in the user s manual but still have symptoms that do not seem to match your test results or if you have questions please contact your healthcare professional Plea...

Page 8: ...Use by Temperature limitation Manufacturer Catalogue number Serial number Control Sufficient for Authorized representative in the European Community Blood glucose test result in mg dL Keep away from s...

Page 9: ...of Meter Symbols Date on the left side User flagging patient number Time on the right side Apply Control Solution AM Before Noon Apply blood PM After Noon Unit Result Message Battery Insert a strip T...

Page 10: ...ion Test 18 About The Control Solutions 19 Running A Control Solution Test 20 Understanding Control Solution Test Results 22 Chapter 3 Testing Your Blood Glucose 24 Using The Lancing Device 24 Inserti...

Page 11: ...38 Viewing And Deleting Test Results 39 Running With Your Computer 40 Chapter 5 Maintenance And Troubleshooting 41 Inserting Batteries 41 Cleaning Your Meter 42 Cleaning Your Lancing Device 42 Mainten...

Page 12: ...oll through results Test Strip Slot Insert test strip here Battery Door Flip open the battery door by pushing the tab in the direction of the arrow and pulling the door up Right Button Press to enter...

Page 13: ...on Area Electrode Lancing Device Trigger button Press the trigger button to activate the lancing device Carrier Insert the lancet into the carrier Adjustable tip Select the desired penetration depth H...

Page 14: ...on the back of the meter by pushing the tab in the direction of the arrow and pulling the door up 2 Insert two batteries The meter will beep to confirm the batteries are inserted correctly 3 Put the...

Page 15: ...ng mode 3 The screen display flashes the last 2 digits of the year Press Left Button or Right Button to adjust the year and press Power Button to confirm the setting 1 When you replace the battery the...

Page 16: ...of is on it allows you to keep testing records of multiple users separate by marking them from 001 999 2 The meter defaults to the medical record function being Press Right Button to reset this funct...

Page 17: ...to check the expiration date on the test strip bottle The test strip will be expired either in six months after the date of the bottle is opened or till the expiration date printed on the bottle Store...

Page 18: ...r meter and test strips are working properly to give reliable results You should run the control solution tests when You think the meter or test strips may be working incorrectly You drop the meter Yo...

Page 19: ...chever comes first Do not use a control solution that is past the expiration date Control solutions can stain clothing If you spill it wash your clothes with soap and water Close the bottle tightly af...

Page 20: ...ace like a table 4 Remove the control solution bottle cap and wipe the tip of the bottle with a tissue 5 Squeeze the bottle until a tiny drop forms at the tip of the bottle and place solution on top o...

Page 21: ...ion on the top of test strip The meter starts to count down from 5 seconds and will show the results 7 Do not remove the test strip until you confirm that the reading falls within the range printed on...

Page 22: ...for the appropriate control solution level Make sure you compare the result to the correct level of control When the control solution result is inside the range on the test strip bottle your test str...

Page 23: ...well You can use the control solutions to verify the meter s functions Is the control solution expired or contaminated If yes replace with a new control solution to check the performance of SMBG syste...

Page 24: ...r you Please do not share your lancing device with anyone And always use a new sterile lancet Lancets are for one time use only Used test strips and lancets are considered bio hazardous waste in accor...

Page 25: ...cet into the lancing device to get it ready for use Lancets are for use only and a new sterile lancet should be used each time you perform a test 1 Unscrew the Cap 2 Insert the lancet into the lancing...

Page 26: ...lancing devices and lancets may transmit blood borne pathogens such as viral hepatitis 6 Set the lancing device aside until later in the test 5 Pull on the sliding barrel of the lancing device until i...

Page 27: ...nning A Blood Glucose Test With Blood From Your Fingertip 4 Place the lancing device against the pad of your finger Press the trigger button to activate the lancing device 1 Wash your hands with soap...

Page 28: ...n strip to make it to the confirmation window Not enough blood on strip to get a test result 5 Gently squeeze and or massage your fingertip until a round drop of blood on your fingertip 6 Touch the bl...

Page 29: ...sh the Strip Ejector to eject the test strip dispose of both strip lancet properly 9 After discarding wash hands thoroughly with soap and water Rinse and dry thoroughly 10 Clean and disinfect the mete...

Page 30: ...ervals In a pre meal or fasting state more than 2 hours since the last meal Two hours or more after taking insulin Two hours or more after exercise Do NOT use AST if You think your blood glucose is lo...

Page 31: ...sult from the alternate site test does not match how you feel do a fingertip test to confirm the result again Do NOT change your treatment just because of an alternate site result do a fingertip test...

Page 32: ...device for AST testing 1 Massage the puncture area of forearm for a few seconds 2 Press and hold the device with clear adjustable tip against the forearm 3 Press the trigger button to activate the la...

Page 33: ...conds 2 Press and hold the device with one clear adjustable tip against the palm 3 Press the trigger button to activate the lancing device 4 Hold the device against palm and increase pressure until th...

Page 34: ...he cap 2 Without touching the used lancet stick the lancet tip into its protective cover 3 Pointing the lancing device toward a container for sharp or biohazard material slide the ejection button down...

Page 35: ...on Standards of medical care in diabetes Diabetes care 2013 Vol 36 Suppl 1 S21 Unusual Test Results If your test result does not match the way you feel please follow these steps 1 Run a control test C...

Page 36: ...3 3 mmol L and you have symptoms of high blood glucose contact your healthcare professional instantly If the result does not match how you feel follow the steps under Unusual Test Results What It Mean...

Page 37: ...drawn from your arm by a healthcare professional Keep in mind that the lab could use different technology than EASYMAX MU Blood Glucose Meter and that blood glucose meters for self testing generally...

Page 38: ...d as the latest is added so it is very important to have the correct time and date set in the meter 1 Do not change your therapy based on one individual result in memory 2 The memory is not lost when...

Page 39: ...s Left Button or Right Button to view the test results of each patient s number 2 The test results of control solution shows without Patient s numbers 3 To delete a test result press Left Button for m...

Page 40: ...be turned on or turned off 1 Use the USB cable to connect to the meter and your computer The display shows PC 2 The meter starts to transfer data to the GlucoManager TM software 3 When all data is up...

Page 41: ...ll not turn on the first time batteries are inserted Please press and hold Power Button or insert the test strip to turn your meter on The meter will turn off automatically Or you can press and hold P...

Page 42: ...oth dampened with water and mild detergent DO NOT places the entire device under water To disinfect the cap after cleaning place it in 70 75 rubbing alcohol for 10 minutes at least once a week Allow t...

Page 43: ...turn it on and lets you know if something is wrong See Screen Messages and what to do about them To make sure the display is working properly turn off the meter Press and hold Power Button to see the...

Page 44: ...local dealer or healthcare professional Message What it means What to do Humidified Used strips The meter has detected a problem with the test strip Repeat the test with a new strip Refer to pages 27...

Page 45: ...his error message Err 001 appears again please contact your local dealer Memory Error Replace the batteries first If ERROR 005 appears again please contact your local dealer No result in memory The te...

Page 46: ...r than 630 mg dL 35 0 mmol L Re check your glucose level If the result is HI again obtain and follow instructions from your healthcare professional without delay The test result is lower than 20 mg dL...

Page 47: ...0 F 104 F The LF and thermometer icon appears Temperature is too low outside the required range of 10 C 40 C 50 F 104 F This alerts users that an incorrect result may occur if the test continues Reloc...

Page 48: ...d transportation condition Temp 20 C 50 C 4 F 122 F Relative Humidity R H 90 Blood sample 0 6 L Fresh blood from fingertip palm or forearm Hematocrit Hct 20 60 Power 2 Alkaline 1 5V AAA Battery life O...

Page 49: ...ct the results Hematocrit below 20 may cause higher results Hematocrit above 60 may cause lower results 7 Altitude up to 3 048 meters above sea level has no effect on readings Healthcare Professionals...

Page 50: ...8 mmol L Ascorbic Acid 4 mg dL 0 26 mmol L 10 89 1 76 4 55 Ibuprofen 50 mg dL 2 43 mmol L 3 10 2 88 4 62 L Dopa 1 8 mg dL 0 09 mmol L 10 59 7 91 4 90 Sodium Salicylate 50 mg dL 3 12 mmol L 2 59 9 42...

Page 51: ...MAX US Office 9550 S Eastern Ave Suite 253 Las Vegas NV 89123 www easymaxdiabetescare com Obelis s a Boulevard G n ral Wahis 53 B 1030 Brussels Belgium TEL 32 2 732 59 54 FAX 32 2 732 60 03 E mail mai...