background image

Quick Start Guide

All-in-One Blood Glucose Meter

Summary of Contents for C50

Page 1: ...Quick Start Guide All in One Blood Glucose Meter ...

Page 2: ...Contents Parts summary Setting the time and date Preparing the meter Performing a test Connecting to the App Clean the Meter E 4 E 5 E 9 E 12 Key to Error Messages ...

Page 3: ...1 Sensor cover 2 Display 3 Sample settings button 4 Memory recall button 5 Penetration depth selector 6 Lancet holder 7 Lancet 8 Lancet ejector Parts summary ...

Page 4: ...9 Test Cassette 10 Thumbwheel 11 Battery cover 12 Lancet cover 13 Penetration depth indicator 14 Sensor ...

Page 5: ...Setting the time and date Hold down the S button to enter the settings menu 1 2 Press M to change the flashing character Press S to save ...

Page 6: ...1 Remove the protective film from the screen Pull protective tab from the battery compartment 2 Preparing the meter Slide the sensor cover to the left as far as it will go Meter will turn on ...

Page 7: ...3 Insert cassette rounded end first and press flat Close sensor cover 4 Slide lancet cover open ...

Page 8: ...5 6 Insert lancet and twist off cap Close lancet cover and set desired penetration depth ...

Page 9: ...Performing a test 1 2 Wash your hands in warm water Dry your hands thoroughly Watch the video ...

Page 10: ...3 4 Open the sensor cover and wait for the wind on symbol Turn the thumbwheel until a test zone is in position and screen displays OK ...

Page 11: ...h selector firmly against selected puncture site then press S Gently squeeze out a small round drop of blood Use immediately Do not allow the drop to run dry or spread To repeat close open the sensor cover to re cock ...

Page 12: ...7 Apply blood Aim for the red flashing light ...

Page 13: ...Cover the red illuminated area with blood or the meter will display E 7 8 You may press down on the blood drop to spread it Do not touch the test zone Results in approx 5 seconds ...

Page 14: ...rn the thumbwheel to dispose of the used test until OK is displayed 9 10 NOTE If you close the cover before winding the used test away E 3 will be displayed Close the sensor cover The meter will turn off ...

Page 15: ...Connecting to the App Hold down the S M buttons until the bluetooth symbol appears 1 2 Open the app on your mobile device and follow the instructions ...

Page 16: ...Clean the Meter E 4 E 5 E 9 E 12 Remove the cassette Push on the curved grip lift out from the top 1 2 Clean sensor window with a damp cotton tip or tissue in the direction shown ...

Page 17: ...our meter in the user manual E 6 Cassette wound too far Test zone wound past sensor window Continue winding test zone into the waste chamber until OK appears on the display Close the sensor cover to turn meter off or continue winding to the next test zone Cassette wound too far when winding used test into waste chamber The cassette has been wound too far causing a fresh test to be withdrawn from t...

Reviews: