Recording
Start recording
Manual Start
Press and hold
Start/Stop
button for
3
seconds to manually active data logger, if a logger is set with
following start conditions, then recording would only start after these conditions are fulfilled.
Start Delay
Start delay time and a
Clock
display on LCD and logger begins countdown
after activated, frequency of countdown refreshes as per delay time
remains, for example, when there is 2 hours left, delay time refreshes each
hour, 30 minutes left, LCD refreshes each 1 minute.
(In this case, start delay time is 30 minutes)
Condition Start (Temperature Start)
LCD displays current temperature reading after logger activated manually,
reading refreshes every 30 seconds. Logger will change to recording status
after temperature condition is met.
Auto Start
Device starts recording on pre-set time automatically.
“REC” appears in upper middle screen indicating logger is currently
recording.