UM-ITAG4-160217 REV.D

 

 

Auto Start 

Device automatically starts recording on the specific date/time preset, no need to press

 Start/Stop

 

button after configuration. 

 

Button Settings 

 

Disable Stop Button 

This option is used to avoid mistaken Stop Button action, once enabled, device could not be 

stopped by pressing 

Start/Stop

 button, recording stops when logging cycle completes, however, 

you are still allowed to stop recording through TempCentre. 

 

Alarm Settings 

 

Alarm Mode 

 

No alarm setting 

 

High & Low alarm, 2 alarm ranges determined by 2 alarm limits. 

 

Multiple Alarm ranges, 5 alarm ranges determined by 5 alarm limits. 

 

High alarm zones: A1, A2, A3   

 

Ideal range: A4 (No alarm will be triggered when data falls within this range ) 

 

Low alarm zone: A5, A6   

 

 

Only ITAG4 Bio supports multiple alarm ranges. 

 

Alarm Type 

 

Single Event 

Alarm could only be triggered when data over high limit (or under low limit) not less than an alarm 

delay continuously.

 

 

Cumulative Alarm 

Alarm could be triggered once the cumulative total time data over high limit (or under low limit) is 

not less than an alarm delay. 

 

 

Summary of Contents for ITAG 4

Page 1: ...2016 Shenzhen EO Technology Co Ltd www eofirm com USER MANUAL EN ITAG 4 ITAG 4 Bio ITAG 4 TH DATA LOGGER UM ITAG4 160217 REV D www eofirm com...

Page 2: ...ng Configuration 8 Configuration Options 8 Recording 11 Start Recording 11 In Recording 12 Stopping recording 14 Downloading Data 15 Downloading Data by Obtaining PDF Report 15 Downloading Data via Te...

Page 3: ...UM ITAG4 160217 REV D PRODUCT OVERVIEW 1 USB Plug 2 Start and Stop Button 3 View Button 4 LCD 5 Humidity sensor window only available on ITAG4 TH...

Page 4: ...r 6 Start timer and limit violation duration indicators 7 Recording indicator 8 Time units 9 Readings 10 Low alarm zone status temperature sensor 11 High alarm zone status temperature sensor 12 Limit...

Page 5: ...RH Accuracy 0 5 C 0 9 F 0 5 C 0 9 F 0 5 C 0 C 60 1 C rest range 4 RH 20 RH 80 RH 5 RH rest range Resolution 0 1 C F Logging Interval 1 minute 2 hours Alarm Range High and low alarm High and low alarm...

Page 6: ...ownloading then an email containing download link would be sent to the mail address you left in couple of minutes If you want to download the software from www eofirm com again after your successful r...

Page 7: ...elect configuration parameters you need then download the configuration profile cfg format file save it into the logger created removable storage device TempSenDisk LCD displays USB when logger connec...

Page 8: ...r describing monitoring task and showing on data report Logging Interval The recording and sampling interval for data logger optional from 1 minute to 2 hours Startup Mode Manual Start Device needs to...

Page 9: ...entre Alarm Settings Alarm Mode No alarm setting High Low alarm 2 alarm ranges determined by 2 alarm limits Multiple Alarm ranges 5 alarm ranges determined by 5 alarm limits High alarm zones A1 A2 A3...

Page 10: ...ogger Time Base Select Time Zone and enable disable Day Light Saving time DST for PDF report Language Select language for PDF report LCD Screens after Configuration Manual Start LCD displays Stby afte...

Page 11: ...ording starts Start Delay Start delay time and a Clock appear on LCD and logger begins countdown after logger started manually refresh frequency of countdown time changes as per remaining delay time f...

Page 12: ...Event Current Reading 12 3 C RH channel only available on ITAG TH No Alarm High alarm zone Temp High alarm zone RH Violation event 2 Violation event 0 Low alarm zone Temp Low alarm zone RH Violation...

Page 13: ...ber on LCD indicating the violation events please refer to data report for the real number of violation events when LCD shows 99 for violation events LCD Examples View Logging Statistics Click on View...

Page 14: ...logging cycle completes memory full LCD displays HoLd while pressing button to stop device to remind user current button action is effective and please keep unit recording stops Recording Stops StoP...

Page 15: ...lowed to open this PDF file with TempCentre for further analysis and archiving and upload it to TempSen s Cold Chain CloudTM online data management analysis and sharing center www coldchaincloud com t...

Page 16: ...emember to save data to data base of TempCentre by clicking Apply button in software DATA RESULTS EXPLANATION Current Status Running status of logger when downloading data Start Time The time the firs...

Page 17: ...ows on LCD and in Events column of alarm statistics in data report Alarm First Triggered Time First Triggered The time an alarm is first triggered at For example logger is set to logging interval of 1...

Page 18: ...power system disabled shelf life validity of device is two years from manufacturing date Expiry date is printed on the back of each logger please make sure you will use the logger before expiry date...

Page 19: ...perature is a simplified way of expressing the overall effect of temperature fluctuations during storage or transit of perishable goods Calculation E R ln 1 2 E activation energy standard value 83 144...

Page 20: ...l status To avoid damaging your data logger please do not disconnect it from USB port while it is communicating with the computer Do not use any aggressive cleaning agents or solvents to clean device...

Page 21: ...unique 10 digits alphanumeric number automatically created for identifying each monitoring task Trip Description A brief note entered by user for describing a monitoring task and shows on data report...

Page 22: ...ified by the TempSen trademark trade name or logo affixed to it The Limited Warranty does not apply to any non TempSen hardware product or any software even if packaged or sold with the TempSen hardwa...

Page 23: ...2016 Added recording results explanation C 06 07 2016 Corrected few mistakes D 05 18 2017 Violation time calculation method changed T 86 755 8420 0058 F 86 2571 2314 E sales eofirm com W www eofirm co...

Reviews: