![envea STACK 602 User Manual Download Page 58](http://html1.mh-extra.com/html/envea/stack-602/stack-602_user-manual_2415516058.webp)
Appendix A
–
Alignment
STACK 602
58
(PC-000848-MA Issue 5, December 2021)
ENVEA
UK Ltd
Figure 27: Alignment principles
–
Receiver.
Alignment Procedure
Aligning the Sensor Heads using the Alignment Tool
An alignment tool (
) is provided that simulates the effect of the lens and
Receiver diode.
1.
Power up the Receiver. Check the Transmitter is ON (producing a green light
source), then install the Receiver (accurate alignment of the sensor head is
not essential at this stage).
2.
Install the alignment tool into the Receiver mounting as follows:
2a.
Undo the grub screws behind the Receiver enclosure (refer to
and remove the complete enclosure from the sensor body.
2b.
Insert the alignment tool with the wide end first into the sensor body.
Figure 28: Alignment tool.
Summary of Contents for STACK 602