9 Maintenance
STACK 602
42
(PC-000848-MA Issue 5, December 2021)
ENVEA
UK Ltd
7.4.1
Sensor Channels
For each STACK 602 sensor, the following channels are added to the control unit
automatically. Switch to the
Multiple Devices
screen to view these.
Channel
Description
1
Dust
Dynamic opacity reading (can be calibrated in mg/m³)
2
Opacity
Opacity reading in % (0% = clear stack, 100% = obscured
stack)
3
I
Total light intensity (diagnostic only)
4
SC
Scintillation count (diagnostic only)
5
Tx Temp
Transmitter PCB temperature (diagnostic only)
6
Rx Temp
Receiver PCB temperature (diagnostic only)
The STACK 602 optical sensor has two outputs:
Dynamic Opacity
mode and
Opacity
mode. The Dynamic Opacity mode is equivalent to a ‘Dust’ reading. If
either mode (602 Dust / 602 Opacity) is not required, this can be deleted manually
from the control unit.
7.5
Adjust Sensor Brightness and Sensitivity
To set up the STACK 602, the brightness and sensitivity settings must be adjusted,
so an appropriate Received Light signal is being detected at the Receiver.
CAUTION!
Adjustments to the sensor sensitivity and/or brightness affect
the Dust calibration.
Neither sensitivity nor brightness settings should be changed once the
STACK 602 has been calibrated. The settings should be noted during
the calibration procedure (see section
1.
below, select the
QA/Self Tests
icon to display this
screen.
2.
From the
Device
list, select
Stack
n
Opacity
.
3.
Select the
Set Up
button (top right) and press ENTER to display this screen.
NOTES
: Sensitivity and brightness settings can be set automatically or manually. It is
recommended to set the brightness and sensitivity automatically using the
AUTOSET function (see below)
.
7.5.1
Adjusting Sensitivity/Brightness Automatically (Autoset)
1.
Autoset
button and press ENTER.
2.
Select the
Sensitivity
button and press ENTER to increase the setting from
Low
to
Medium
/
High
until a sufficiently large Received Light signal is
reached (
>12.5
, on the bar chart).