![envea PicoFlow Operating Instructions Manual Download Page 32](http://html1.mh-extra.com/html/envea/picoflow/picoflow_operating-instructions-manual_2415514032.webp)
32
7.
Digital input
Selection of function for external control.
7.1
Digital input 1
Submenu
7.1.1 Function
Selection:
None:
digital imput disabled
none / reset totaliser /
Reset totaliser:
totaliser would be
AutoCal
reset to zero
AutoCal:
an auto calibration will be start
7.1.2 Working direction
Selection:
NO / NC
If necessary, invert the value of the input
level.
7.1.3 Filter
Input:
0.1 … 99.9 s
Time during which the requested signal must
remain pending.
7.2
Digital input 2
Same as digital input 1