2. Calibration
Deposit a calibration curve
Depending on the selection under
8.2.4 Calibration
, the parameters to be entered are changing.
Average-Calibration
From the average value of all sensors, a common calibration table is created for throughput
calculation.
2.1
Calibration factor
Input:
0.01 … 9.99
Factor for the subsequent adjustment of the
actual measurement. All measurements are
scaled by this factor.
2.2
Calibration filter [s]
Input:
1 … 9999
Filter time for recording the raw value during
calibration. It would be made an average out
of the measured RAW-values.
2.3
Calibration points
Input:
2 … 5
Number of support points for
a linearisation above the operating range.
2.4 Calibration
Calibration sub-menu
2.4.1 P1 value
Input:
measurement
Output measurement in the selected
mass/time unit.
2.4.2 P1 calibration
Transfer:
raw value
Transfer of the current raw value (filtered)
from the mass flow with the key [
←
].
The value can also be entered directly.
. . . (depending on the number of support points)
For additional support points (depending on
[ 2.3 ]), additional value pairs can be set.
2.4.n Pn value
Input:
measurement
2.4.n Pn calibration
Transfer:
raw value