20
The edited data is transmitted to the evaluation unit via “Device program”.
Critical data concerning the ModBus communication and the calibration must be confirmed before the
parameters are transmitted to the evaluation unit:
✔
If, when saving the the parameters in the evaluation unit, the system calibration data
is changed, this action must be confirmed by checking “Overwrite calibration”.
✔
If, when saving the the parameters in the evaluation unit, the system interface parameters
are changed, this must be confirmed by checking the selection “Overwrite baud/addr.”.
In addition, with the PC software,
•
the evaluation unit parameters can be saved in a file (save configuration)
•
the evaluation unit parameters can be loaded from a file (load configuration)
•
the evaluation unit parameter can be printed via the windows standard printer (print configuration)
•
the measured values can be logged in a data logger file (enter the file name and storage rate, and
activate the data logger in the On-Line representation)
The software language can be set by right-clicking the “Sprache/Language/Langue” field in the bottom
program line on “Deutsch/English/Français”.
Protection against unauthorised use:
The PC interface does not have a password prompt as it is assumed that only authorised personnel will have
access to the PC and the software. However, the password to operate the display can be read and changed
in menu
8. System
in
8.6 Password
.