2C.
ELECTRICAL/GAS CONNECTION
The electrical connection of the fume cupboard ECOCHEM is carried out by connecting the power cable
located on the upper of the right side of the cabinet to a suitable power point (see technical table). When
the cabinet is connected, the green light on the control panel switches on (see section 4D).
If provided by the laws in force, insert an automatic protection switch (at its maximum power) upstream on
the power input line. The switch shall be fitted with a differential relay with operating current rate not
exceeding 30 mA.
Any instrument with a heat source shall be located at a distance of at least 150 mm from the bottom and
side walls. Moreover we recommend not to use a flame at less than 300 mm from the adsorbing filter.
2D.
POSITIONING OF THE CABINET ON THE SUPPORTING TABLE
ECOCHEM cupboards can be supplied with the relevant supporting table, which they have to be fixed to.
After assembling the supporting table (see instructions at section. 14) put the supporting table on a flat non
sloping floor and be sure that all the parts of the cabinet, that can be opened (control board panel and
window) are locked. Then position the cabinet on the table.