8C.
PREFILTER REPLACEMENT
(by the user)
-
Open the front control panel, rotate and unlock the tightening lock (13), then lower the pre-filter case
(6).
-
Remove the prefilter from the relevant case
-
Insert the new prefilter in its seat.
-
Close the prefilter case (6) and fix it with the tightening lock (13)
8D.
ACTIVATED CHARCOAL FILTER REPLACEMENT
(By the User)
ATTENTION:
For safety and environmental reasons, the worn-out activated charcoal filter/s shall be
handled with rubber gloves and deposited in polyethylene bags
-
Open the front control panel (1) unscrewing the fastening socket head screws (2) and block it with the
correct bracket (3) located on the right side of the panel.
-
Raise the activated charcoal filters (4) by unscrewing the correct knobs (5) for a few centimetres.
-
Remove the activated charcoal filter (4) from its slide guide.
-
Insert the new activated charcoal filter/s (4) into the relevant slide guides so that it correctly leans on
the structure.
-
Turn the knobs (5) of the filter group and block the activated charcoal filter (4).
-
Close the control panel (1) and fix it with the correct screws.
-
When you switch on the fume cupboard again, set the installation date (only for the "M" version).
8E.
FLUORESCENT LAMPS REPLACEMENT
(by the user)
-
Open the front control panel (1) unscrewing the fastening socket head screws (2)
-
And block it with the correct bracket (3) located on the right side of the panel.
-
Turn the fluorescent tube (8) by 90° and remove it.
-
Put the new fluorescent tube (8) into the lamp holder and turn it of 90°.
-
Close the front control panel (1) and fix it with the correct screws.