INVUE GV148 LIQUID CHEMICAL CONCENTRATION MONITOR
19
User Guide | Entegris, Inc.
– Results of the Rezero Self Test will also be
posted to the Alarm Log viewable on the
“Alarms” tab
Analog Outputs
This
Self Test
sets the analog outputs either high or
low so that their output could be checked with an
external measuring device.
1. To test the analog output, set it to low or high and
press “Start”.
2. Measure the output with appropriate device during
this 30 second cycle.
3. At the end of the 30 seconds, the outputs will revert
to represent their normal values.
TOTAL INTERNAL REFLECTION (TIR) TAB
This tab allows you to view 2 TIR curves at a time.
To load new TIR curves, press Load New for both
TIR A and TIR B. Notice that you can now hide
one or both curves.
Press Load File to display a saved or reference
TIR curve for either A or B, or Save File to store
on your PC