Page 18 of 56
EN109-1498740-40 IFU
5
Installation
Inspection
In case of damage from transport is noticed, contact your local distributor.
DO NOT USE the device!
Immediately upon unpacking the device, perform the following steps:
•
Verify the delivery documents to make sure that the delivery is complete.
•
Verify that the packaging contains all the items listed in the standard accessories list
(Refer to chapter 4 - Package Contents in this manual).
•
Check the external components and accessories for possible damage due to transport.
Connection to mains supply
Use of any other cable other than the supplied cable is strictly PROHIBITED as it
affects patient safety and the proper function of the device.
Before connection of this unit to the mains supply, check that the mains voltage and
frequency stated on the identification plate of this apparatus corresponds with that of the
mains supply.
Make certain that the device is electrically grounded by connecting only to a grounded
electrical service receptacle conforming to the applicable national and local electrical codes.
Do not place the device in a location where the power cord could be tripped over or pulled
out during treatment!
Plug in the device to a wall socket using the mains cable.
•
Insert the mains cable into socket and connect it to a wall socket.
•
Set mains switch to On.
•
Power LED indicator is lit green indicating that the device is connected to the mains supply.
•
Turn on the device with On / Off button.
•
The device will initialize and perform a self-test. This may take a while.
•
At the end of the self-test the device enters the Home menu and is ready for use.
Placing the optional battery
Do not interchange the black and red wires as this will damage your device!
The battery contains material that is noxious to the environment. Observe the local
regulations when disposing of the battery!
Due to the high current demand of electrotherapy applications, we recommend to explicitly
use batteries supplied by Enraf-Nonius B.V. (part number 2501.016).
•
Remove the mains cable from the mains connector.
•
Place the Endomed 484 device upside down and on a soft surface.
•
Remove the two screws from the battery cover using a screwdriver.
•
Slide and lift the battery cover.
•
Align the battery on the bottom of the main unit with the polarity of battery terminals in the
correct position. The polarity is marked at the bottom of the battery compartment.
•
Locate the black wire and attach it to the
–
terminal of the battery.
•
Locate the red wire and attach it to the + terminal of the battery.
•
Slide the battery upside down into the battery compartment taking care that the wires do not
get jammed.
•
Place and slide the battery cover back into position.
•
Secure the battery cover with the two screws using a screwdriver.
•
Place the device back on its feet.
•
Reconnect the mains cable to the power line connector.