Energist UltraPlus VPL™ System
User Manual
SD1-43 Issue 3 (CCF 296)
9
1.8 Technical Specification
Light Source:
Single Flashlamp, Reflective Cavity, Filter
Wavelength Range:
530nm Handpiece: 530nm – 950nm
610nm Handpiece: 610nm – 950nm
Acne Handpiece: 405/630-950nm
Handpiece Waveguide:
BK7 Glass waveguide
Handpiece Filter:
530nm/610nm - Cut-on Absorption filter
Acne - Dichroic filter
Max. Pulse Energy:
255 Joules
Pulse Duration:
3*ms, 5*ms and 7*ms
Pulse Sequencing:
2 – 15 pulses
Pulse Delays:
1*ms – 20*ms
Pulse Train Times:
7*ms – 385*ms
Repetition Rate:
Up to 0.5 Hz
Treatment area:
5 cm
2
(5cm x 1cm)
Energy Density:
530nm Handpiece: up to 51 J/cm
2
610nm Handpiece: up to 45 J/cm
2
Acne Handpiece: up to 25 J/cm
2
(Energy densities accurate to ±20%)
Interface Control Panel:
Full Colour LCD display with Touchscreen.
Windows XP
Electrical:
100V (6.5A) to 240V (3.0A), 60/50 Hz.
Continuous mode of operation.
Cooling:
Self contained, de-ionised water cooled
Environmental:
10
O
C to 30
O
C Non-condensing humidities below
75% RH.
NOT
to be used in explosive, oxygen-
enriched or nitrous oxide atmospheres.
Dimensions:
950mm x 430mm x 420mm (H x W x D)
Weight:
Approximately 54kg
Classification:
Electrical Class I, Type BF, Mobile Equipment.
Class 2b Rule 9 – 93/42/EEC. Risk Group 2.
Ingress Rating:
IPX0. The system is not protected against the
ingress of liquids and should be used only in a
clean, dry environment.
*Pulse timing is microprocessor controlled to ± 2% accuracy