Energist UltraPlus VPL™ System
User Manual
SD1-43 Issue 3 (CCF 296)
20
1. Open side panel (middle right) where handpiece connections are
located
2. Remove main cable from holder
3. Depress water connector tab and gently remove. Shake to dispel
excess water
4. Pull on outer sheath of power/communications cable connector to
release lock and pull firmly out
4.4 Changing the Applicator
If the system is
ON
, the applicator may be changed by:
1. Select
Change
below the
Applicator information
(top right hand side)
2. Tap
Yes
to confirm
3. Wait for system to confirm it is safe to change the applicator
4. Open side panel (middle right) where handpiece connections are
located
5. Depress water connector tab and gently remove. Shake to dispel
excess water
6. Pull on outer sheath of power/communications cable connector to
release lock and pull firmly out