Energist UltraPlus VPL™ System
User Manual
SD1-43 Issue 3 (CCF 296)
123
Guidance and manufacturer’s declaration – electromagnetic immunity
UltraPlus is intended for use in the electromagnetic environment specified below. The
customer or the user of the system should assure that it is used in such an environment.
Immunity test
IEC 60601 test level
Compliance
level
Electromagnetic environment –
guidance
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
3Vrms
150kHz to 80MHz
3V/m
80MHz to 2.5GHz
3Vrms
3V/m
Portable and mobile RF
communications equipment
should be used no closer to any
part of the UltraPlus, including
cables, than the recommended
separation distance calculated
from the equation applicable to
the frequency of the transmitter.
Recommended separation
distance:
P
d
2
.
1
=
P
d
2
.
1
=
80M to 800MHz
P
d
3
.
2
=
800M to 2.5GHz
where
P
is the maximum output
power rating of the transmitter in
Watts [W] according to the
transmitter manufacturer and
d
is
the recommended separation
distance in meters [m].
Field strengths from fixed RF
transmitters, as determined by an
electromagnetic site survey
a
,
should be less than the
compliance level in each
frequency range
b
.
Interference may occur in the
vicinity of equipment marked with
the following symbol:
NOTE 1: At 80MHz and 800MHz, the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is
affected by absorption and reflection from structures, objects and people.
a
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless)
telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV
broadcast cannot be predicted theoretically with accuracy. To assess the
electromagnetic environment due to fixed RF transmitters, an electromagnetic site
survey should be considered. If the measured field strength in the location in which the
UltraPlus is used exceeds the applicable RF compliance level above, the
UltraPlus should be observed to verify normal operation. If abnormal performance is
observed, additional measures may be necessary, such as reorienting or relocating the
UltraPlus system.
b
Over
the
frequency range 150kHz to 80MHz, field strengths should be less than 3V/m.