6 Commissioning
TopClean S CPC 30
66
EHauser
Fig. 28: Example for the programme sequence of "Clean C" and "Clean Int".
C07-CPC300xx-05-06-00-de-008.EPS
Clean C
Hold
Hold and after-effect time
Prog.
Cleaning / calibration
Cleaning / calibration
Cleaning / calibration
Clean Int
Hold
Hold and after-effect time
Hold and after-effect time
Measuring time 1800 s
Measuring
time
CODE
CHOICE
(default = bold)
INFO
(E1, 2 = Editor types, s. page 28)
User
settings
N1
Configuration
Activate
Selection
Configuration
= Create/edit a TopClean S programme
Activation
= Switch on/off TopClean S functions
E1
Configuration:
NA1
Weekly progr.
Clean trigger
Ext. control
off
off
off
Note in display:
Displays the current system status
E2
!
Note!
Fields NA2 to NA4 can only be edited if all the functions in Field NA1 are switched off and the option to control an additional valve is available.
NA2
Valve V1
------
Assignment of additional valves (if available)
You can assign different functions to the additional
valve:
Sterilisation, Sealing water or user-defined (user
programme).
Sealing water
: If you have activated the sealing water
function, you can use it with every assembly move (see
user programme, Field NAD1).
!
Note!
• If you change the function of the additional valve,
sterilisation or sealing water is no longer available in
the set programmes.
• If you have changed the function, you must check
the assignment in the user programmes.
E2
NA3
Valve
(0...9; A...Z)
Enter valve name
Enter 8-character name for valve .
E2