TopClean S CPC 30
5 Operation
Hauser
29
5.1.7
Factory setting
All the factory settings are active when the instrument is switched on for the first time.
The table below lists all the main settings. For all further factory settings, please refer to
the "Description of functions" chapter (from p. 34); the factory settings are printed in
bold
.
5.2
Replaceable memory
The DAT module is a memory device (EEPROM) which is plugged into the connection
compartment of the transmitter. Using the DAT module you can
•
save
the complete settings, the logbooks and the data logger of a transmitter and
•
copy
the complete settings to other CPM 153 transmitters with identical hardware
functionality.
This considerably reduces the effort to install or service several measuring points.
Parameter
One-circuit instrument
Select operating mode
pH
Select measuring principle
One-circuit circuit 1
Select electrode type
Glass electrode 7.0
Select connection type
Solution ground
Select temperature display
Deg. C
Select temperature compensation
ATC
Temperature measurement
Off
Select temperature sensor
Pt 100
Contact functions
NAMUR
Select current output 1
pH/redox K1
Select current output 2
Temperature K1
Hold
Active with PARAM and CAL
(inactive with DIAG)
Current
output 1:
0/4 mA value:
20 mA value:
pH 2 / –1500 mV / 0.0% / 0.0 °C
pH 12 / +1500 mV / 100.0% /
100.0 °C
Current
output 2:
0/4 mA value:
20 mA value:
Temperature
0.0 °C
100.0 °C